JOG NOW : Service & maintenance manual - rev. 6.2
Pagina 11.4
11.4.
LIST OF CRITICAL SPARE PARTS
Technogym medical devices are identified by “Serial N” plates that allow them to be tracked
throughout the useful life-span of the device. Just as for products, there is also a list of so-called
“critical” components which, through the Serial N, must be guaranteed as traceable.
The list of these critical components is shown in the table below
:
Code
Component
0WR00233XX
Elevation motor
0WR00500XX
Tread belt motor
W0004467XX
AT MD power supply box
0WK00429XX
Display Board 500
0WQ00348XX
C-Safe Board
0WR00633XX
HR Cardio Receiver
0WCU0243XX
Power Supply cable (LED with AT-UL)
W0006892XX
Power Supply Box 16A MED
-
Display Program –SW version
Note:
XX = Baan Revision
YY = Language
If a critical component listed in this table is replaced during a repair, maintenance or refurbishment,
this fact must be
noted
in the
SERVICE REPORT
by recording the
tracking information
for the
new component.
Содержание Jog Now Excite+ 700
Страница 2: ......
Страница 4: ...JOG NOW Excite Service maintenance manual rev 6 2...
Страница 29: ...JOG NOW Excite Service maintenance manual rev 6 2 Page 2 17 UNITY Version...
Страница 31: ...JOG NOW Excite Service maintenance manual rev 6 2 Page 2 19 UNITY Version...
Страница 186: ...JOG NOW Excite Service maintenance manual rev 6 2 Page 8 6 Page intentionally left blank...
Страница 222: ...JOG NOW Excite Service maintenance manual rev 6 2 Pagina 12 2 Blank page...
Страница 224: ...JOG NOW Excite Service maintenance manual rev 6 2 Pagina 13 2 Blank page...
Страница 226: ......
Страница 227: ......