Edition 1.0.1 March / 2020
38/42
TechnoAP Co., Ltd.
(3)
Measurement of Energy spectrum
When measuring the spectrum, make the following settings on the "config" tab, then click the "Config" menu. Set the
"threshold" value kept in the waveform measurement to "threshold" in the "config" tab.
Config tab
Open the "spectrum" tab and check the settings shown below, then click the menu "Clear" → "Start". After execution,
the following spectrum will be displayed.
Energy spectrum measurement
Please note the following points.
1. Check CH 1 of "spectrum on / off" so that CH 1 spectrum can be displayed.
2. When analyzing peaks, set ROI. For details, refer to "5.5.spectrum tab".
The measured data can be saved in the menu "File" - "save histogram".
To end the measurement, click the menu "Stop".