Issue/Rev. 0.0 (5/19) ║ MNIN003 • Page 15

Installation 

Operation 

Manual 

      

Detector Calibration – 4528-X A with 4-20mA Module Mounted "Piggy Back" on  

Detector Card

1.   A.   Calibration 4528-X A With 4-20 mA Module Mounted “Piggy Back” on Detector Card
 

1.   Remove Detector aluminum face plate by removing the three 6-32 mounting screws and the bottom screw  

 

 

on P.C. Card.

 

2.   Mount the P.C. Card in the condulet by using the short screws removed from the Detector assembly. Be sure  

 

 

the upper left screw at 10 oʼclock is tight as this screw “grounds” the P.C. Card to the condulet. Retain all 

 

 

 

screws for future use.

 

3.   Connect 24 V.D.C. to the left terminal block of the detector. Observe polarity. Press clip (on probe wire) onto  

 

 

the stud coming into condulet from probe.

 

4.   Turn on power/system to cause oil to flow through the probe. Allow flow to stabilize. For calibration 

 

 

purposes the oil should preferably have less than 1% water in it.

 

5.   Set Probe/Ref. switch at top of chassis to probe position.

 

6.   Set a DC volt meter to 0-10 volt scale. Insert leads into test jacks on detector face.

 

7.   Turn the Fine Zero (marked ZERO F) clockwise (CW) 25 turns until the pot is fully CW. Full clockwise 

 

 

is indicated by a “clicking feel and sound when the pot is turned.

 

8.   Turn the fine zero back counter clockwise (CCW) approximately 10 turns.

 

9.   Turn the coarse zero pot (marked ZERO C) 25 turns full counter clockwise. Listen for clicks.

 

10. Turn the coarse zero CW until the volt meter reads 0.00 to 0.05. Do not turn pot past point where 

 

 

numbers stop getting smaller.

 

11.  Next, turn the fine zero to get as small a reading possible on volt meter.

 

12. Move Probe/Ref. switch to Ref. position.

 

13. Adjust output span control to make volt meter show 5.00 volts.

 

14. Move Probe/Ref. switch to Probe position.

 

15. Remove volt meter leads from test sockets. Turn volt meter to milliampere setting and connect leads 

 

 

to “signal” terminals.

 

16. Adjust the single, small blue pot (R1) on piggy-back module to get reading of 4 mA on meter.

 

17. Reinstall Detector face plate using the long 6-32 screws at the 10 oʼclock and 2 oʼclock positions and

 

 

 

the short screws at the middle and bottom.

 

18. Remove volt meter leads from signal terminals. Turn volt meter back to voltage setting and insert 

 

 

leads into red and black sockets.

 

19. Run a “shake-out” of oil going through probe to determine amount of water in it.

 

20. On a 5% Detector turn COARSE/FINE tuning adjustments as required to make voltmeter read the 

 

 

water cut.

 

21. On a 10% Detector turn COARSE/FINE tuning adjustments as required to make voltmeter read 1/2 the 

 

 

water cut.

 

22. On a 20% Detector turn COARSE/FINE tuning adjustments as required to make voltmeter read 1/4 the 

 

 

water cut.

 

23. Remove volt meter.

 

24. Connect field wiring as required to the signal terminals for 4-20 mA output.

Содержание 4528

Страница 1: ...Water Cut Monitor Bulletin MNIN003 Issue Rev 0 0 5 19 Installation Operation Manual 4528 Detector CX 645 Probe CSA Approved 4728 Monitor MANUAL ...

Страница 2: ...ation Supervision Start Up and Commissioning Services Available Customer Support Contact Information Customer Service TechnipFMC FMC Technologies Measurement Solutions Inc 1602 Wagner Avenue Erie Pennsylvania 16510 USA P 1 814 898 5000 F 1 814 899 8927 measurement solutions TechnipFMC com TechnipFMC com Literature Library http info smithmeter com literature online_index html ...

Страница 3: ... 10 External Powered 10 Typical Applications 12 Typical Tank Flow Diagram 13 Detector Calibration 4528 0 5 VDE Output 14 Detector Calibration 4528 X A with 4 20mA Module Mounted Piggy Back on Detector Card 15 Detector Information Type B and C 17 Installation Hints 17 Troubleshooting 18 Recommended Spare Parts 18 4728 with Explosion proof Enclosure Dimensional Drawing 19 4728 with Weather proof Enc...

Страница 4: ...Page 4 MNIN003 Issue Rev 0 0 5 19 Page intentionally left blank ...

Страница 5: ...easured by the system Alarm relays and a 4 20 mA output can be used to connect the system to other supervisory or control elements Specifications Input Power 115 VAC 15 50 60 Hz 2 watts 220 VAC Optional 14 28 Volts DC 130 mA S W Range 0 5 0 10 or 0 20 S W Alarm Setpoint 0 100 of S W Range Input Signal 0 5V DC from 4528 Detector Indicators S W Setpoint LCD Merchantable Oil LED Green Excess S W LED ...

Страница 6: ...30 sec on 3 sec off Analog Output The analog output is 4 20 milliamperes corresponding to 0 5 0 10 0 20 depending on unit range The 4 20 mA output function is accomplished by the use of a precision integrated circuit The 4 mA point may be adjusted slightly with the zero control on the Model 4728 The 20 mA point is factory preset The Model 4728 utilizes a loop switch to determine whether the 4 20 m...

Страница 7: ...1 which is located at the lower right of the P C card The position of this switch determines whether the 4 20 mA output is to be internally powered from the 4728 or externally powered from an outside source Care must be taken to be sure this switch is in its proper position BEFORE an external 4 20 mA loop power source is connected Failure to have the switch in the proper position may cause compone...

Страница 8: ... can be accomplished by removing the three screws marked A in the above draw ing and gently lifting the top plate off of the printed circuit board Find the short white or red wire with the clip on it and gently pull the wire straight up Remove the nut from screw B Insert the printed circuit board into the condulet Position the board so the rod sticking up from the probe goes into the round hole on...

Страница 9: ...ng the three screws marked A on the drawing Wiring Diagram Note Spade Lugs are not used on CSA Approved Model 4728 Note Typical Interconnecting cable between Detector and Readout Maximum recommended length is 300 feet Installation Operation Manual ...

Страница 10: ...al 4 20 mA Loop Warning Loop Switch MUST be up in external position BEFORE connecting external power on 4728 with bottom terminal strip Loop Switch MUST be up before connecting external power on CSA Approved Model 4728 with the two sides and bottom terminal strips ...

Страница 11: ... capacitance Nominal Pipe Size Catalog Model Number Stock Number 2 3 4 6 CX 645 200 BGP CX 645 300 BGP CX 645 400 BGP CX 645 600 BGP 81001190 81001195 81001200 81001205 2 3 4 6 CX 645 200 BFP CX 645 300 BFP CX 645 400 BFP CX 645 600 BFP 81001191 81001196 81001201 81001206 2 3 4 6 CX 645 230 BGP CX 645 330 BGP CX 645 430 BGP CX 645 630 BGP 81001192 81001197 81001202 81001207 2 3 4 6 CX 645 260 BGP ...

Страница 12: ...he process Percent Water Recording In this application the Indicating Unit provides local reading on an indicating meter and provides an analog output signal to a standard recorder In place of the recorder this output could be connected to a meter computer input or any other DC input device Percent Water Indication and Set Point Control In this application the adjustable alarm may be used to opera...

Страница 13: ...Issue Rev 0 0 5 19 MNIN003 Page 13 Installation Operation Manual Typical Tank Flow Diagram ...

Страница 14: ... pot is turned 4 Turn the fine zero back counter clockwise CCW approximately 10 turns 5 Turn the coarse zero pot marked ZERO C 25 turns full counter clockwise Listen for clicks 6 Turn the coarse zero CW until the reading is at or near the grindout value Do not tune through zero Always approach the grindout value from the CW direction 7 Next turn the fine zero to get complete agreement of the displ...

Страница 15: ...imately 10 turns 9 Turn the coarse zero pot marked ZERO C 25 turns full counter clockwise Listen for clicks 10 Turn the coarse zero CW until the volt meter reads 0 00 to 0 05 Do not turn pot past point where numbers stop getting smaller 11 Next turn the fine zero to get as small a reading possible on volt meter 12 Move Probe Ref switch to Ref position 13 Adjust output span control to make volt met...

Страница 16: ... supply a 4 20 mA signal direct to an external readout device It is not designed to be used with a Model 4728 Readout Unit 3 Maximum Loop Resistance is 250 ohms 4 Shielded twisted pair wire is recommended for the 4 20 mA output Ground the shield at one end of the wire only preferably at the Readoutʼ end 5 Starting in mid year 2000 the 4 mA pot adjusting screw will be accessible through a hole just...

Страница 17: ...roximately 2 below zero Open Probe Chassis Pickup wire is not connected to the probe Shorted Probe Occurs when electrical short exists in the probe Installation Hints Always install probe up stream of dump valve anytime probe is mounted on oil line coming off treater separator etc Never install downstream as gas will break out to some degree as oil goes from vessel pressure to lower line pressure ...

Страница 18: ...I gravity oils reading high est A digital volt meter set on low DC volts range should read less than 25 volts A reading of any voltage indicates the probe insulators are becoming contaminated and should be changed Normally a monitor will operate in a normal manner up to 25 volts but then the readings will become erratic Solution to problem is elimination of coating buildup causes or periodic clean...

Страница 19: ...Issue Rev 0 0 5 19 MNIN003 Page 19 Installation Operation Manual 4728 with Explosion proof Enclosure Dimensional Drawing ...

Страница 20: ...Page 20 MNIN003 Issue Rev 0 0 5 19 Installation Operation Manual 4728 with Weather proof Enclosure Dimensional Drawing ...

Страница 21: ...Issue Rev 0 0 5 19 MNIN003 Page 21 Installation Operation Manual CX 645 Probe Dimensional Drawing ...

Страница 22: ...ny P 49 4101 304 0 The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are currently in effect Otherwise the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect Contact Information Field Service Response Center 24 ...

Отзывы: