INTENDED USE 

 

Fully Automatic Electronic Sphygmomanometer is for use by medical professionals or at home 
and is a non-invasive blood pressure measurement system intended to measure the diastolic 
and systolic blood pressures and pulse rate of an adult individual by using a non-invasive 
technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference 
is limited to

 

22cm-48cm(approx. 8 21/3

2″

~18 29/32

). 

 

CONTRAINDICATION 

 

  It is inappropriate for people with serious arrhythmia to use this Electronic 

Sphygmomanometer. 

 

PRODUCT DESCRIPTION 

 

Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure 
and pulse rate can be measured automatically and non-invasively. The LCD display will show 
blood pressure and pulse rate.The most recent 60 measurements can be stored in the 
memory with date and time stamp.The Electronic Sphygmomanometers corresponds to the 
below standards: IEC 60601-1:2005/EN 60601-1:2006/AC:2010 (Medical electrical equipment 
-- Part 1: General requirements for basic safety and essential performance), 
IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment -- Part 1-2: 
General requirements for basic safety and essential performance - Collateral standard: 
Electromagnetic compatibility - Requirements and tests), EN 1060-1: 1995 + A1: 2002 + A2: 
2009 (Non-invasive sphygmomanometers - Part 1: General requirements), EN 1060-3: 1997 + 
A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part 3: Supplementary 
requirements for electro-mechanical blood pressure measuring systems). 
 

SPECIFICATIONS 

 

1.  Product name: Blood Pressure Monitor 
2. Model: 

TMA-6000 

3.  Classification: Internally powered, Type BF applied part,IPX0,No AP or APG,Continuous 

operation 

4.  Machine size: Approx. 154mm×104mm×45mm (6 1/16” x 4 3/32” x 1 25/32”) 
5.  Cuff circumference: 22cm-30cm(8 21/32

-11 13/16

), 30cm-42cm(11 13/16

-16 17/32

) (Optional), 42cm-48cm(16 17/32

-18 29/32

) (Optional) 

6.  Weight: Approx. 234g (8 1/4oz.) (exclude batteries) 
7.  Measuring method: Oscillometric method, automatic inflation and measurement 
8.  Memory volume: 60 times with time and date stamp 

9.  Power source: DC 6V 600mA, batteries: 4 ×1.5V

SIZE AA   

10. Measurement range: 

Cuff pressure: 0-300mmHg 
Systolic:     60-260mmHg 

Содержание TMA-6000

Страница 1: ...TIONS 3 NOTICE 4 SETUP AND OPERATING PROCEDURES 5 1 BATTERY LOADING 5 2 CLOCK AND DATE ADJUSTMENT 6 3 CONNECTING THE CUFF TO THE MONITOR 6 4 APPLYING THE CUFF 7 5 BODY POSTURE DURING MEASUREMENT 7 6 T...

Страница 2: ...usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are awake and ac...

Страница 3: ...lectrical equipment Part 1 General requirements for basic safety and essential performance IEC60601 1 2 2007 EN 60601 1 2 2007 AC 2010 Medical electrical equipment Part 1 2 General requirements for ba...

Страница 4: ...o allow the blood circulation in your arm to recover Prolonged over inflation cuff pressure exceed 300 mmHg or maintained above15 mmHg for longer than 3 minutes of the bladder may cause ecchymoma of y...

Страница 5: ...nd complied with IEC 60601 1 EN 60601 1 UL 60601 1 and IEC 60601 1 2 EN 60601 1 2 UL 60601 1 2 is suitable for this monitor Please note that the monitor jack size hole 3 5mm center pin 1 0mm Please pa...

Страница 6: ...ode press the START and MEM button simultaneously for two seconds the month will blink at first Press the button START repeatedly the day hour and minute will blink in turn While the number is blinkin...

Страница 7: ...PECIFICATIONS to make sure that the appropriate cuff is used 2 Measuring on same arm each time 3 Do not move your arm body or the monitor and do not move the rubber tube during measurement 4 Stay quie...

Страница 8: ...rtbeat symbol if any will blink on the screen The result will be automatically stored in the monitor d After measurement the monitor will turn off automatically after 1 minute of no operation Alternat...

Страница 9: ...ntended to provide a basis of any type of rush toward emergency conditions diagnosis based on the color scheme and that the color scheme is meant only to discriminate between the different levels of b...

Страница 10: ...ened Apply the cuff correctly and try again Body posture was not correct during testing Review the body posture and testing sections of the instructions and re test LCD Display shows abnormal result S...

Страница 11: ...chnical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measu...

Страница 12: ...magnetic emissions The TMA 6000 is intended for use in the electromagnetic environment specified below The customer or the user of the TMA 6000 should assure that it is used in such an environment Emi...

Страница 13: ...requency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Tabl...

Страница 14: ...tions Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordles...

Страница 15: ...istance according to frequency of transmitter m Rated maximum output power of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3...

Отзывы: