Control  valves  or  stop  cocks  should  always  be  installed  downstream  from  the  sensor  to 
assure that the pipe is always full of liquid. 

3.3 

Mixtures 

If  liquids  of  different  conductivities  are  mixed  it  is  necessary  to  install  the  sensor  a 
minimum of 50 x DN from the point of mixture in order to obtain a uniform conductivity of 
the liquid and stabilize the readings.
If this distance is less, the readings may be unstable.   

    50 DN 

3.5 

Pumps 

Pumps should be mounted upstream from the sensor to avoid the suction part of the pump 
(vacuum).

5

Содержание FLOMAT FX FlomatFX

Страница 1: ...FLOMAT FX Electromagnetic insertion sensor Series Instructions Manual R MI FlomatFX Rev 1 English version instrumentation for fluids...

Страница 2: ...4 MOUNTING THE INSERTION FITTING 7 5 MOUNTING THE SENSOR 9 5 1 Tightening torque 9 5 2 Electronic converter connection 9 6 CONFIGURATION 9 6 1 Fc factor 9 6 2 Pipe diameter 9 7 POSSIBLE PROBLEMS WHEN...

Страница 3: ...k the instrument removing any packaging material that could be attached to the sensor Do not remove any grease from the collar that fits the sensor to the electronics housing 2 2 Storage temperature 2...

Страница 4: ...sensor must be BS 1042 2 2 1983 standard 3 1 Sensor position The most adequate position is in the side of the pipe In this way deposits of particles on the electrodes and air pockets at the top of the...

Страница 5: ...different conductivities are mixed it is necessary to install the sensor a minimum of 50 x DN from the point of mixture in order to obtain a uniform conductivity of the liquid and stabilize the readi...

Страница 6: ...G in the range of 20 150 Hz according to IEC 068 2 34 3 7 Magnetic fields Strong magnetic fields close to the sensor should be avoided 3 8 Temperature In open air installations it is recommended to i...

Страница 7: ...e thickness Peel off the bottom part of the label Once the insert is placed into its final position where the label was cut must coincide with the exterior of the pipe This ensures that the measuring...

Страница 8: ...us example if the gasket was 5 mm thick the distance that the fitting should extend above the exterior wall would be H 55 3 5 53 mm DN Sensor length Fitting length C H 80 101 93 10 88 s 100 12 5 85 5...

Страница 9: ...In the event that during the installation of the sensor the electronics or the sensor cable were disconnected simply replace the two connectors on the sensor push the mesh above the head tight the si...

Страница 10: ...7 2 1 Integration time It is the time during which the average value is calculated In the supposition that the instrument takes 10 readings per second if an integration time of 5 seconds is selected...

Страница 11: ...housing EN 1 4404 AISI 316L PVDF Electrodes head PVDF Gasket NBR 9 2 Insert fitting connection G2 1 4 thread DIN flange special flange 9 3 General characteristics Accuracy 3 5 with respect to measure...

Страница 12: ...due to misuse being excluded all responsibility due to any other damage or the effects of wear caused by the normal use of the devices Any consignment of devices for repair must observe a procedure w...

Отзывы: