background image

3

WORKING PRINCIPLE

The Flomat electromagnetic insertion flowmeters are based on Faraday’s induction law. 

When an electrically conductive liquid flows through a magnetic field, perpendicular to the 
flow direction, it induces a voltage E, proportional to the liquid velocity. 

Two  electrodes  in  contact  with  the  liquid  and  positioned  perpendicular  to  the  magnetic 
field, capture this voltage E. 

RECEPTION

The Flomat electromagnetic flowmeters are supplied ready for their installation and 
operation.

They are also supplied packaged for their protection during storage and haulage.  

All the flowmeters have been tested in our calibration rigs to obtain the Fc factor for each 
sensor. For more information about the Fc factor, see point 6.1, page 9 of this manual. 

2.1 

Unpacking

Carefully unpack the instrument, removing any packaging material that could be attached 
to  the  sensor.  Do  not  remove  any  grease  from  the  collar  that  fits  the  sensor  to  the 
electronics housing. 

2.2 

Storage temperature

-20ºC ...... +60ºC  

2.3 

Handling 

This should always be done with care and without knocks. 

INSTALLATION 

This should be made in a point that guarantees that the pipe is always completely full and 
where  a  turbulent  flow  profile  is  completely  developed  (follow  recommendations  in  point 
3.2)

Avoid  high  points  of  the  pipes  where  air  pockets  usually  form,  or  pipes  with  falling  flow 
where it can form a vacuum. 

Partially full pipes can produce important reading errors.  

If  the  point  where  the  Flomat  should  be  installed  has  a  discharge  point  near  to  it,  it  is 
necessary to install a siphon in order to avoid any stagnant air in the sensor, as shown in 
the following figure.

E  =  B·v·d 

Where:

E = Measured voltage in the 

electrodes

B = Magnetic flux density 
v  = Average liquid velocity 
d = Distance between electrodes

Содержание FLOMAT FX FlomatFX

Страница 1: ...FLOMAT FX Electromagnetic insertion sensor Series Instructions Manual R MI FlomatFX Rev 1 English version instrumentation for fluids...

Страница 2: ...4 MOUNTING THE INSERTION FITTING 7 5 MOUNTING THE SENSOR 9 5 1 Tightening torque 9 5 2 Electronic converter connection 9 6 CONFIGURATION 9 6 1 Fc factor 9 6 2 Pipe diameter 9 7 POSSIBLE PROBLEMS WHEN...

Страница 3: ...k the instrument removing any packaging material that could be attached to the sensor Do not remove any grease from the collar that fits the sensor to the electronics housing 2 2 Storage temperature 2...

Страница 4: ...sensor must be BS 1042 2 2 1983 standard 3 1 Sensor position The most adequate position is in the side of the pipe In this way deposits of particles on the electrodes and air pockets at the top of the...

Страница 5: ...different conductivities are mixed it is necessary to install the sensor a minimum of 50 x DN from the point of mixture in order to obtain a uniform conductivity of the liquid and stabilize the readi...

Страница 6: ...G in the range of 20 150 Hz according to IEC 068 2 34 3 7 Magnetic fields Strong magnetic fields close to the sensor should be avoided 3 8 Temperature In open air installations it is recommended to i...

Страница 7: ...e thickness Peel off the bottom part of the label Once the insert is placed into its final position where the label was cut must coincide with the exterior of the pipe This ensures that the measuring...

Страница 8: ...us example if the gasket was 5 mm thick the distance that the fitting should extend above the exterior wall would be H 55 3 5 53 mm DN Sensor length Fitting length C H 80 101 93 10 88 s 100 12 5 85 5...

Страница 9: ...In the event that during the installation of the sensor the electronics or the sensor cable were disconnected simply replace the two connectors on the sensor push the mesh above the head tight the si...

Страница 10: ...7 2 1 Integration time It is the time during which the average value is calculated In the supposition that the instrument takes 10 readings per second if an integration time of 5 seconds is selected...

Страница 11: ...housing EN 1 4404 AISI 316L PVDF Electrodes head PVDF Gasket NBR 9 2 Insert fitting connection G2 1 4 thread DIN flange special flange 9 3 General characteristics Accuracy 3 5 with respect to measure...

Страница 12: ...due to misuse being excluded all responsibility due to any other damage or the effects of wear caused by the normal use of the devices Any consignment of devices for repair must observe a procedure w...

Отзывы: