8.6.1.COVOL 

If the Pulse input is configured for Covol (jumpers in position 2-3), a potential free 
contact can be connected between terminals 1 & 2. Terminal 1 is connected to the 
common of the instrument. This input has a 1 M

 pull-up resistor connected to 

terminal 2 and to the 3,3 V power supply. The input has a hysteresis with the 
switching points situated at about 1 V & 2,1 V. If an open collector transistor is to 
be used as the switching element instead of a COVOL, then the aforementioned 
parameters must be taken into account. 

Due to the presence of a filter to avoid the effects of contact bounce, the maximum 
pulse frequency for the COVOL input is 300 Hz.  
The minimum input frequency is 0.06 Hz  

   

8.6.2 .Turbine

 

If the Pulse input is configured for Turbine (jumpers in position 1-2), a coil of a 
magnetic pick-up can be connected between terminals 1 & 2. 
The input impedance is 1,2 K

. This input is protected to limit the maximun applied 

voltage to the circuit. This protection consists of two diodes in parallel and two 100 
Ohm resistors.   
The maximum voltage that can be applied to the input is 10 Vpp. Voltages higher 
that this can cause damage to the equipment. 
The maximum input frequency is 5000  Hz 
The minimum input voltage is   7  mVpp 

 8.7. Loop connection 

The 4-20 mA loop is connected to a screw terminal block marked as 

mA

. The right hand 

side terminal is the positive (+) and the left hand side terminal is the negative (-). 

Host or 
Power Supply

CH420

+

+

-

-

  

             Complies with the Directive EMC 89/336/EEC

 

1 M

2

2

1

1

+3.3V

100

1K

100

2

2

1

3

Pick-up

Содержание CP420R

Страница 1: ...Programmable flow rate transmitter CP420R CH420R Instructions Manual R MI CH420R Rev 0 English version...

Страница 2: ...There are three push buttons with the following functions 2 1 Programming In all the programming screens the RESET push button exits the screen without saving the data in the memory even if changes in...

Страница 3: ...rogramming In the first screen we program the flow rate equivalent to 4 mA lower range The measuring units will be the ones chosen in the previous screen The push buttons have the same function as the...

Страница 4: ...in the instrument Example 1 Shortage of volume If we have a flow meter body which specifies i l 1 985 and when we check the volume of a batch we find that instead of having 100 litres as programmed we...

Страница 5: ...er of Device Variables 2 Number of Dynamic Variables 1 Mappable Dynamic Variables No Number of Common Practice Commands 13 Number of Device Specific Commands 2 Bits of Additional Device Status 12 Alte...

Страница 6: ...digits 7 Size of the digit 8 mm Reset By means of push button 8 3 Flow rate Indication N of digits 5 Size of the digit 5 mm 8 4 General characteristics Ingress protection IP40 Ambient temperature rang...

Страница 7: ...imum input frequency is 0 06 Hz 8 6 2 Turbine If the Pulse input is configured for Turbine jumpers in position 1 2 a coil of a magnetic pick up can be connected between terminals 1 2 The input impedan...

Страница 8: ...specified in the order the result of service or modification by un authorized persons bad handling or accident This warranty is limited to cover the repair or replacement defective parts which have no...

Отзывы: