MODELS 

__  420  __ 

__   

 

 

 

 

C  COVOL input 

 

 

 

 

T  Turbine input 

 

 

 

L  Local mounting on the flow meter 

 

 

 

R  DIN rail mounting inside a switchboard 

 

P  Transmitter with analog output 

 

H  Transmitter with analog output and HART communication 

 

 

 

8.

TECHNICAL CHARACTERISTICS 

8.1. Power  supply 

2 wires, by means of the loop current. 
The instrument has a protection diode to avoid damage if the power supply is connected 
with inverted polarity . 
 
Nominal voltage:                             7.5 ... 36 Vdc

 

Power consumption:                       

 20 mA 

8.2. Totalizer 

N. of digits:                                      7

 

Size of the digit:                              8 mm 

Reset:                                              By means of push button  

8.3. Flow rate Indication 

N. of digits:                                      5

 

Size of the digit:                              5 mm 

8.4. General characteristics 

Ingress protection:                          IP40

 

Ambient temperature range:           0 … +60 ºC 

8.5. Input terminals 

The pulse input is connected to a screw terminal block marked as 

Pulse

Terminal n.                        COVOL                      Turbine 
         1                               Common                        Live 
         2                                   Live                            Live 
           

8.6. Inputs characteristics 

 
NOTE: The input terminals are not isolated from the 4-20 mA loop. Under no 
circumstances can an electrical connection be made between the 4-20 mA loop and 
the inputs. In the event of supplying various transmitters with the same power 
supply one must make sure that the inputs isolated between each other.  
 

Содержание CP420R

Страница 1: ...Programmable flow rate transmitter CP420R CH420R Instructions Manual R MI CH420R Rev 0 English version...

Страница 2: ...There are three push buttons with the following functions 2 1 Programming In all the programming screens the RESET push button exits the screen without saving the data in the memory even if changes in...

Страница 3: ...rogramming In the first screen we program the flow rate equivalent to 4 mA lower range The measuring units will be the ones chosen in the previous screen The push buttons have the same function as the...

Страница 4: ...in the instrument Example 1 Shortage of volume If we have a flow meter body which specifies i l 1 985 and when we check the volume of a batch we find that instead of having 100 litres as programmed we...

Страница 5: ...er of Device Variables 2 Number of Dynamic Variables 1 Mappable Dynamic Variables No Number of Common Practice Commands 13 Number of Device Specific Commands 2 Bits of Additional Device Status 12 Alte...

Страница 6: ...digits 7 Size of the digit 8 mm Reset By means of push button 8 3 Flow rate Indication N of digits 5 Size of the digit 5 mm 8 4 General characteristics Ingress protection IP40 Ambient temperature rang...

Страница 7: ...imum input frequency is 0 06 Hz 8 6 2 Turbine If the Pulse input is configured for Turbine jumpers in position 1 2 a coil of a magnetic pick up can be connected between terminals 1 2 The input impedan...

Страница 8: ...specified in the order the result of service or modification by un authorized persons bad handling or accident This warranty is limited to cover the repair or replacement defective parts which have no...

Отзывы: