3 - Technical Data
Emissions
Tecan Vacuum Separator (Te-VacS™) Operating Manual, 391236, en, V4.0
19 / 54
3.6.2 Chemical Resistance to Vapors
(1) resistant, (2) resistant to a limited extend, (3) not resistant
Parts
DMSO
Acetonitrile
Methanol
Methylene
Chloride
Hexane,
Chloroform
DMF, Ethyl
Acetate
Ethanol,
Isopropanol
Sodium
Hypochlorite, 10%
Waste bottle
sealing
2
1
1
3
3
2
1
1
Vacuum con-
trol valve:
Sealing,
O‑rings
1
3
1
2
1
3
1
1
Pressure sen-
sor
3
3
2
3
3
3
1
2
Table 3:
Parts with contact to vapors
3.6.3 Foam Formation
Substances causing foam formation should be avoided. Please note that often
protein solutions can be a source of foam.
If foam formation cannot be avoided, it is recommended to install a second waste
bottle with sufficient tubing.
Refer to
“Connecting an Additional Waste Bottle” [
.
3.7 Emissions
Refer to the Operating Manual of the instrument. Refer to section
.
3.8 Software Requirements
The Te-VacS may be operated by the following software packages:
•
Freedom EVOware
•
FluentControl
•
Other Tecan software packages supporting Te-VacS
It is strongly recommended to use the latest software version. For further
information please consult the