4 - Putting into Operation
Setting up the Balance Kit and the SAG 285/01
Balance Kit Operating Manual, 393278, en, V1.0
4 - 5
Computer with
USB Ports (no
RS-232)
Modern computers, particularly notebooks, often have no RS-232 ports at all, but
are equipped with two or more USB-ports. In such a case the instrument and the
Control Unit must be connected via appropriate USB to Serial converters
according to the following figure.
Note:
The connection of the computer to the instrument via an USB to Serial
converter will reduce the communication speed between the instrument and the
computer. As a consequence, the execution time for the Gravimetric Test may
increase by 10 to 20 %, compared to the time it takes when the instrument is
connected to the computer directly.
Recommended converters: See sections
and
Fig. 4-4
Connection via USB to Serial converter
USB
Cnv
RS-232
RS-232
RS-232
Optibo-Board
Notebook
Instrument
H
K
L
J
I
M
H
Weighing Cell
I
Cable Weighing Cell to Control Unit
J
Control Unit
K
LC-RS9 (Control Unit to computer)
L
Power supply for SAG 285/01
M
Cable instrument to computer
Cnv
USB to serial converter