6 - 2
Balance Kit Operating Manual, 393278, en, V1.0
6 - Troubleshooting
Troubleshooting Table
1 USB, 1 RS-232 port:
Connections mixed
up.
Connect instrument to RS-232 port and
Control Unit to USB.
Balance Setup func-
tion in Setup & Ser-
vice SW carried out
incorrectly
Repeat Balance Setup and test if Control
Unit communicates properly with Instrument
Software.
During Gravimetric Test
the LiHa is not moved to
the correct position above
the Weighing Cell
Grid position of
adapter plate does not
correspond to workta-
ble map
Place the adapter plate in the correct grid
position.
Wrong worktable map
loaded
Load the correct worktable map.
Tab. 6-1
Troubleshooting table (cont.)
Problem / error
Possible cause
Corrective measures