![Teac TD-9000T Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/teac/td-9000t/td-9000t_instructions-for-use-manual_3983500006.webp)
6
2. Communication functions
2-1. FIELD NETWORK
2-1-1. SETTINGS
To open the FIELD NETWORK screen, touch the buttons in the
following order on the Home Screen.
e
CALIB. & SYSTEM
e
SYSTEM
e
FIELD NETWORK
CONTROL INPUT
Select the device control method.
Only one method can be used for control.
Control terminal: Use external contact.
COMM. (Communication control): Use communication.
MEMORY CONTROL
Select the sensor number and work number switching
method.
FORBID: Use control input signals.
ALLOW: Use communication.
(Work switching is only enabled when work number
switching is set to external input in the work settings.)
EtherNet/IP
Ethernet settings
2-2. EtherNet/IP
Refer to the following values for connection input/output
settings.
Input settings
Instance ID: 100
Data size: 48-byte
Output settings
Instance ID: 150
Data size: 2-byte
2-2-1. Settings
To open the EtherNet/IP screen, touch the buttons in the fol-
lowing order on the Home Screen.
e
CALIB. & SYSTEM
e
SYSTEM
e
FIELD NETWORK
e
SETTINGS (next to “EtherNET/IP”)
The input items change according to the DHCP setting.
When DHCP is off
Set the following items.
IP ADDRESS
SUBNET MASK
When DHCP is on
Values for the following items will be set by DHCP. They cannot
be input.
IP ADDRESS
SUBNET MASK
MAC
The MAC address of this unit will be shown.