12
Calibration
Content
Command
Command No. Byte
Parameter
Note
Sensor
Sensor memory
1000
1
0: Sensor Memory 1, 1: Sensor
Memory 2, 2: Sensor Memory 3,
3: Sensor Memory 4
Sampling
1006
1
0: 5 kHz, 1: 25 kHz
Y axis
1007
1
0: Load, 1: Load and displacement
X axis
1008
1
0: Time, 1: Displacement
X axis full scale
1009
1
0:80 ms, 1: 170 ms, 2: 400 ms, 3: 800 ms,
4: 2.0 s, 5: 4.0 s, 6: 10.0 s, 7: 30.0 s,
8: 60.0 s, 9: 90.0 s
The definitions
change according
to the X axis setting
(time/displacement).
Unit shown setting
1401
1
0: None, 1: N, 2: kN, 3: kPa, 4: MPa, 5: g,
6: kg, 7: ton, 8: mNm, 9: Nm, 10: kNm,
11: dN, 12: Pa, 13: mBar, 14: Bar, 15: m/s
2
,
16: Gal, 17: mm
Maximum display value
(MAX. DISP. VALUE)
1404
4
1 to 99999
Sensor input logic
1405
1
0: Standard, 1: Reversed
Low-pass filter
2001
1
0: OFF, 1: 3 Hz, 2: 10 Hz, 3: 30 Hz,
4: 100 Hz, 5: 300 Hz, 6: 1000 Hz
Moving average number
(MOVING AVG. NUM.)
2002
2
0: Disabled, 2 to 2048: Moving average
number
Automatic digital filter
(AUTO DIGITAL FILTER)
2003
1
0: OFF, 1: ON
Digital zero limit (D/Z
LIMIT SETTING)
2302
4
00000 to 99999
Digital offset
2303
4
±19999
Load cell
Bridge voltage
1001
1
0: 2.5 V, 1: 5 V, 2: 10 V
Load decimal point
position
1002
1
0: None, 1: 0.0, 2: 0.00 3: 0.000, 4: 0.0000
Zero point input
calibration
1003
4
−3100 to +3100, (0.000 to 3.100 mV/V)
Zero balancing
1004
1
1: Execute
Remote sensing
1005
1
0: Unused, 1: Used
Reset zero balancing
1100
1
1: Execute
Rated output value
(RATED OUTPUT)
1101
4
100 to 3200, (0.100 to 3.200 mV/V)
Rated capacity value
(equivalent input)
1102
4
1 to 99999
Rated capacity value
(actual load)
1103
4
1 to 99999
TEDS calibration (TEDS
CALIB.)
1104
1
1: Execute
2. Communication functions