To open the standard click stimulus
fi
le, click
File
, click
Open
, browse to:
C:\TDT\BioSigRZ\SIG\Click.sig
Then click
Open
.
On the main toolbar, click
Modify
, and then click
Signal
.
In the Signal Parameters dialog box, enter the voltage and the corresponding dB level in
the Calibration area, and then click
OK
.
To save the change, click
File
then
Save
.
This will save the change to the stimulus used by BioSigRZ's standard click con
fi
guration.
Run the stimulus
fi
le in BioSigRZ
without
a calibration (.tcf)
fi
le applied to the stimulus. Your
stimulus is already calibrated, and adding a .tcf
fi
le will cause the calibration to be incorrect.
b.
c.
d.
5.
6.