TDT BH32 Скачать руководство пользователя страница 9

System 3

17-25

BH32 Behavioral Cage Controller

To Change the Host name (NetBIOS name):

Type the desired host name in the 

Host Name

 box and click the 

Save 

Config

 button.

Note: 

The Host name can be no greater than 15 characters long and cannot 

contain spaces or the following characters: \ / : * ? " ; | -

Direct

 

Connection

 

to

 

a

 

PC

The BH32 interface can be connected directly to a PC or laptop; however, it is 

usually necessary to use an Ethernet crossover cable to connect the devices. Once 

connected, several steps are required for the PC to recognize the UDP interface 

connection. This method may be performed on any operating system which supports 

TCP/IP. 

To initialize the PC for a direct connection in Windows XP:

1. Physically connect the BH32 Ethernet interface and the PC via an Ethernet 

crossover cable.

2. Click 

Start | Control Panel

 then double-click 

Network Connections

3. Right-click the desired connection (this is usually a Local Area Connection) 

and select 

Properties

4. Select 

Internet Protocol (TCP/IP)

 or if there are multiples, select 

Internet 

Protocol (TCP/IPv4).

5. Click the 

Properties

 button.

6. Select 

Use the following IP address

 and enter these values:

IP address: 

10.1.0.x, where x can be any value from 1 to 254 except 100 

or 101

Subnet mask: 

255.255.255.0

Default gateway: 

Leave empty

7. Click 

OK

The BH32 interface connection should now be recognized by the PC. Cycle power 

on the BH32 device, the IP address of the BH32 will be 10.1.0.101.

Содержание BH32

Страница 1: ...ameter Standard Molex Pin and Socket connectors on the top panel For pinouts see BH32 Technical Specifications on page 17 30 Power The BH32 logic board is powered by either a 6 9V 3A center negative a...

Страница 2: ...and current state of all bits in each bank 3 NetBIOS name and IP address 4 IP address of paired RZ device if paired See RZ Configuration on page 17 22 Push and hold the Mode button for two seconds th...

Страница 3: ...Controlling the Molex outputs with UDP Serial banks A B only Configure the bank as an output The Molex outputs are low when active The digital output logic is the inverse of the Molex output logic Mo...

Страница 4: ...sses configuring the BH32 networking communication and hardware interfaces See RZ UDP Communications Interface on page 1 51 for more information on the basics of networking and the various protocols I...

Страница 5: ...er such as Internet Explorer Chrome or Firefox Enter the device s IP address as the web address e g http 10 1 0 100 and press Enter or Enter the NetBIOS name as the web address e g TDT_BHC_32_1001 and...

Страница 6: ...ogin information can be changed on the Network Configuration page To access a page that requires authentication 1 Click the navigation link for the page 2 When prompted enter the username and password...

Страница 7: ...Device Number is used to identify the BH32 among a network of BH32s For single BH32 use this value should be set to 1 Direction Each Bank can be configured as either Input or Output Logic Level Each...

Страница 8: ...and password can be changed on this page If the Enable DHCP check box is checked the IP Address Gateway Address Subnet Mask and DNS server address values are overridden and automatically configured b...

Страница 9: ...which supports TCP IP To initialize the PC for a direct connection in Windows XP 1 Physically connect the BH32 Ethernet interface and the PC via an Ethernet crossover cable 2 Click Start Control Panel...

Страница 10: ...lly sends a new packet if any of the Inputs change value The 32 bit integer represents the current state of all I O on the BH32 device The ordering is as follows Bit31 Bit0 A8 A1 B8 B1 C8 C1 D8 D1 Exa...

Страница 11: ...his writing This header is used by the BH32 and the RZ to identify packets that they should read and process The target device is identified with the next 24 bits of the packet It consists of a 16 bit...

Страница 12: ...M mode the BH32 display flashes a prompt for the user to press the input button to set the BH32 device number to the starting device number Message Device Group 0b0 0b0000001 0x00000000 Starting Devic...

Страница 13: ...to 256 timestamps total for all pins can be stored in memory Message Device Group 0b0 0b0000110 Pin Number Active 0 or 1 32 bits Reply Device Group 0b1 0b0000110 Pin Number 64 bit timestamp s GET_SET...

Страница 14: ...the BH32 will no longer be in RZ_CONTROLLER state and broaast a GET_SET_TRIGGER message on the network with Trigger Mask 0 Message Device Group 0b0 0b0001101 0x00000000 RZ IP Address optional Reply D...

Страница 15: ...ion 1 C1 Bank C Bits 1 3 5 and 7 14 C2 Bank C Bits 2 4 6 and 8 2 C3 15 C4 3 C5 16 C6 4 C7 17 C8 5 GND Digital I O Ground 18 A1 Bank A Bits 1 3 5 and 7 6 A2 Bank A Bits 2 4 6 and 8 19 A3 7 A4 20 A5 8 A...

Страница 16: ...s are shorted together do not connect a second external power source I O CONNECTORS V Positive Voltage S Signal G Ground Pin Name Description Pin Name Description 1 A1 Bank A Bits 1 3 5 and 7 14 D7 Ba...

Отзывы: