25 |
P a g e
G e n e r a l D a t a H e a l t h c a r e , I n c .
Problem
Description
Remedy
·
The section is broken or the
specimen is torn. (cont.)
·
The blade is uneven
·
Sectioning in the even edge of
the blade or change the blade
·
The section splits or there is a
scratch in length orientation
·
The blade has debris on it.
·
The specimen is scratched by
hard particles
·
Clean the blade.
·
Filtrate the paraffin and pump
it slowly
·
The section sticks to the blade
·
The blade edge has debris.
·
Clean the blade
·
The blade produces scraping
sounds when sectioning, and
the sections are scratched and
show vibration marks
·
The angle of blade is not
proper
·
Clean the blade
·
The blade produces scraping
sounds when sectioning, and
the sections are scratched and
show vibration marks
·
The angle of blade is not
proper
·
Reinstall the blade and adjust
the angle of the blade holder.
·
The surface of the section is
wavy.
·
The angle of blade is not
proper
·
Readjust the angle of the blade
holder.
If there are any other maintenance or service issues that cannot be resolved, please contact TBS-A
Division of General Data Healthcare.