22 |
P a g e
G e n e r a l D a t a H e a l t h c a r e , I n c .
Cleaning the Blade Holder:
•
Remove all the parts of the knife holder as it is showed
in the picture and then clean all the parts separately.
Remember to clean the blade clamp every time before
installing the blade to ensure good sectioning and no
cross contamination.
To obtain a high quality section, it is important to keep the instrument clean. The user
must clean the instrument regularly according to recommendations of the manual to
consistently obtain the best sectioning quality.
Only authorized and qualified service personnel may access the internal
components of the instrument for cleaning and maintenance.
Makes sure the unit is off and disconnected from the outlet before starting
to clean the instrument.
The blade must be removed before taking off the blade holder for
cleaning.
Lock the handwheel before cleaning.
Do Not use any solvents for cleaning the unit.
Make sure no liquid enters the interior of the unit when cleaning.