Seite 8 von 45
General comments
The unit confirms activities acoustically. This fact will not be mentioned in the detailed
description of the functions. Sound and display light can be switched ON and OFF.
The power supply warms up slightly during operation.
Connect the gold plated handheld electrodes – or other application parts - with the BNC-
connectors of the application cable. Place the electrode on the red lead (anode) in the
user’s left hand and the one on the blue lead (cathode) in his right hand, or apply other
electrodes where you see fit (NEVER on an open wound!).
NOTE: The term SQUARE WAVE is used in this document instead of
RECTANGULAR WAVE which is more commonly used in Europe.
How to start the F-SCAN2
The unit is ready for use as soon as
the power supply is connected to an outlet, the
plug attached to the socket POWER and the ON OFF switch on the rear panel set
to ON.
This ‘Start-up’ window is displayed
after power ON. It shows the unit’s
name, the company and country of
origin and the release level of the
internal software.
The window offers the 3 tabs
THERA, DIRP and SETUP to touch
in order to prepare the device for an
application.