Preparat ion:
Connect the EA P-cable to the sensor port of the F-SCA N3.
Connect the EA P – Stylus to the red lead of the EA P-cable.
Connect a handheld or f oot electrode w ith the blue lead of the EAP-cable.
Start the F-SCA N3, open the menu “123“, touch the icon “DIRP“, th en
touch the icon “ EAP“
in the low er right corner of the menu.
EA P- functional test
Touch the tip of the EA P – stylus to the handheld or foot electrode to
cause an electrical short. The measured value must change to 99
accompanied by an audible signal, if this is activated. The value 99 w ill be
stored in the first line LY (lymph point) under the left hand. The cursor-
circle moves to the second line LU (lung). The value 99 must be touched
to re move it and to set the cursor-circle back to line one aw aiting a real
measured value.