TAU STERIL Tau Clave 3000 Скачать руководство пользователя страница 8

 

TAU STERIL s.n.c 

Via Gorizia, 6/a 

Fino Mornasco ( CO ) 

 

 

Instructions for Use 

TAU CLAVE 3000 Fractionated Vacuum 

Data: 30.03.15 

Pag. 8 di 14 

Rev. 3 

N° 1 

 

 

This document may not be reproduced in partial form without the authorization of TAU STERIL. 

 

 

 

 

 

. LED Alarm (red) 

 

Report any anomalies: 

 

they are:   DOOR                      = door open 
                LOW H20 LEVEL    = low level H2o 
                ALARM                   = generic alarm identified by the code on the display time 

 

 

The front panel also has a cup connector (optional), suitable for the attachment of an external printer. It also has a switch with 
fuses (2x10A. 6.3x32).

 

 

5.3  Display mode alarm 

 

Each alarm is displayed in the machine in the following way: 
-

 

 The time display shows the alarm code (ex: A01, A02 etc.). 

-

 

 It has issued a long beep followed by 10 short, eventually silenced. 

-

 

 In case of alarm A00 (door open) or A01 (insufficient water level) which are the most frequent, also they flash their 
LEDS and DOOR LOW LEVEL H20. 

-

 

 In other cases, the ALARM LED flashes and the cause of the alarm must be obtained from the code on the display, or if 
present, from the printer that prints in detail the reason for alarm. 

-

 

The machine remains in this state until the occurrence of the condition of the termination / acknowledgment, specified 
individually for each alarm.

 

-      If this (alarm) occurs during the cycle, the latter is interrupted and the pressure discharged. 

 

5.3.1  Alarms and codes Summary 
 

CODE ALARM 

INDICATIONS 

A00 

Door open: occurs if the start of the cycle, the door is open. Disappears by 
pressing any key or by closing the door.

 

 
 

A01 

Low water: occurs if at the start of the cycle, the water level in the cargo 
tank is insufficient. Disappears by pressing any key or by resetting the 
level. If during the refueling operations is peaked, the time display shows 
"STOP" and the beeper sounds in the manner of an alarm. 2 ceased 
beeper "STOP" disappears, and returns to its normal state.

 

A02 

Tank full exhaust: occurs when the level of water in the recovery tank is 
excessive. Disappears just emptying the tank.

 

A03 

Black-Out: occurs in cases of abandoned cycle by Black-Out. 
Disappears opening the door or pressing any key. * 

A04-A05 

Long time for the heating is energized with any key. * 

A06 

High pressure: it excites with any key. * 

A07

 

High temperature: energizes with any key. * 

A08 

Abnormal pressure: energizes with any key. * 

A09 

Abnormal temperature: energizes with any key. * 

A10 

Stage vacuum not successful. * 

A12 

High temperature PTV: it disappears by pressing any push button * 

A13 

High temperature PTC: it disappears by pressing any push button * 

        

* = Repeat cycle, the material to be sterilized is not sterile. If the problem persists, contact 
            manufacturer. 
 

N.B: A00, A01, A02 do not cause (in the presence of printer) no print activity. 

Содержание Tau Clave 3000

Страница 1: ... TAU CLAVE 3000 Fractionated Vacuum Data 30 03 15 Pag 1 di 14 Rev 3 N 1 1 This document may not be reproduced in partial form without the authorization of TAU STERIL Instructions for Use Tau Clave 3000 Fractionated Vacuum 0426 DIRECTIVE 93 42 EEC and S M I MEDICAL DEVICES ...

Страница 2: ...ning and closure 4 3 Installation 5 4 Characteristics 6 4 1 Technical data 6 4 2 Environmental working conditions 6 4 3 Recommended times and characteristics of programmes 6 5 Operating modes 7 5 1 General information 7 5 2 Visualizations 7 5 3 Alarms visualization modes 8 5 3 1 Alarms and corresponding summary codes 8 5 4 Operating instructions 9 5 5 Description of Tau Clave 3000 Fr Vacuum cycle ...

Страница 3: ... CEI EN ISO 14971 The autoclave Tau 3000 Fractionated Vacuum complies with RoHS 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment 2 2 Expected use of materials to sterilize The equipment is used to sterilize surgical instruments in the medical paramedical sector The materials ...

Страница 4: ...O Instructions for Use TAU CLAVE 3000 Fractionated Vacuum Data 30 03 15 Pag 4 di 14 Rev 3 N 1 4 This document may not be reproduced in partial form without the authorization of TAU STERIL 2 4 Door opening and closure OPENING MOVEMENT CLOSING MOVEMENT ...

Страница 5: ... under the support legs front it is important not to touch the shims On the rear side of the machine is located in a pipe for the steam to exit IMPORTANT The tube must be positioned as shown in Figure 2 Fig 3 Fig 4 Fill the bottle with 2 5 liters of demineralized water then close the bottle with the cap see Fig 4 Introduce the cannula orifice load autoclave Unscrew the tap until the water begins t...

Страница 6: ... with phases of ventilation To select the cycle BOWIE DICK TEST as follows Place the program cycle HOLLOW before viewing 5 sec Of the cycle has finished press the button decreases the time Automatically displays the cycle 134 3 5 of BOWIE DICK TEST PROGRAM FRACTIONATED VACUUM STERILIZATION DRYING AND VENTILATION VACUUM FINAL UNWRAPPED Instruments unwrapped YES 134 4 Minutes 2 11 Bar 15 Minutes YES...

Страница 7: ...se the time in the program Special 6 INCREASE TIME to increase the time in the program Special 5 2 Views 1 PRESSURE It indicates the pressure in bar 2 TEMPERATURE Shows the temperature of steam in the chamber or by pressing the right button TEMP shows the temperature of the heating element Also pressing the left button TEMP you can see the selected temperature cycle 3 TIME Indicating the steriliza...

Страница 8: ...s alarm occurs during the cycle the latter is interrupted and the pressure discharged 5 3 1 Alarms and codes Summary CODE ALARM INDICATIONS A00 Door open occurs if the start of the cycle the door is open Disappears by pressing any key or by closing the door A01 Low water occurs if at the start of the cycle the water level in the cargo tank is insufficient Disappears by pressing any key or by reset...

Страница 9: ...pecial the contents of the display temperature will flash to indicate that the sterilization temperature can be chosen between 105 134 with the temperature The same applies to the time with the time These variations can be carried out simultaneously in both the time for which the temperature 7 Close the door after you have done so 8 Press Start starts the selected cycle 9 Keep area within the open...

Страница 10: ...sufficient ventilation Avoid placing the autoclave near sources of steam or splashing with water that may damage the internal electronics Do not install in places with poor ventilation Do not place it near heat sources 7 Maintenance and Cleaning Each type of maintenance must be carried out with power wire off and the machine completely cold WARNING Do not modify the equipment without the authoriza...

Страница 11: ...e at the side of the probe inside the sterilization chamber see fig 5 Fig 5 Dismantling With the cold machine turn from left to right the steel tube with the help of pliers if necessary paying attention to the probe closer Remove the piece from the chamber unscrew the filter clean it with vinegar or if too dirty replace it with a new one If you want you can avoid disassembling the tube with filter...

Страница 12: ...maximum load 14 Floating common load connection 15 Micro switch door 16 Electric valve N A Normally open 2 Way 24V DC 17 Output optional 24V cc 18 Electric valve N C 2 Way 24V DC 19 Electric valve N C 2 Way 24V DC 20 230 V power supply feeding the vacuum pump and the electric valve nc 2 Vie 21 Bimetallic security 22 Power supply 230 V 50 Hz 23 Resistance 230V 1250 Watts 24 switch 25 Fuse 6 3x32 qu...

Страница 13: ...and electrical equipment or put them out of order It s forbidden the operator to carry out repairs of any kind During cleaning and or maintenance it is absolutely essential to ensure that the equipment is disconnected from the power supply and it is completely cooled In the case of sterilization of surgical instruments eg scalpel or pointed instruments we recommend the use of safety shoes Periodic...

Страница 14: ...Rev 3 N 1 14 This document may not be reproduced in partial form without the authorization of TAU STERIL 10 2 Explanation of Symbols WARNING READ THE INSTRUCTIONS WARNING DANGER BURNS CAUTION RISK OF ELECTRIC SHOCK WARNING USE GLOVES FOR HAND PROTECTION PRODUCT NOT COMPARABLE TO MUNICIPAL WASTE FOLLOW THE RULES FOR MUNICIPAL DISPOSAL ...

Отзывы: