It is always advisable to clean the instruments before putting it in the autoclave. Always clean the chamber (the machine is
cold) with vinegar, then rinse with water and alcohol.
To safeguard against dirt components of the machine, it was put a small filter (unscrewable) on the tube at the side of the
probe inside the sterilization chamber (see fig. 5).
Fig. 5
Dismantling: With the cold machine, turn from left to right the steel tube (with the help of pliers if necessary), paying
attention to the probe closer.
Remove the piece from the chamber, unscrew the filter, clean it with vinegar or if too dirty
replace it with a new one.
If you want you can avoid disassembling the tube with filter, pouring (machine cold) vinegar directly on the filter. To rinse
well with water.
Reassembly: repeat the contrary operations previously carried out, paying attention to the probe.
IMPORTANT: when you place the tube with the new filter, make sure that it is well established, keeping a distance
from the probe equal to 2.5 cm.