TAU STERIL s.n.c
Via Gorizia, 6/a
Fino Mornasco ( CO )
Instructions for Use
TAU CLAVE 3000 Fractionated Vacuum
Data: 30.03.15
Pag. 12 di 14
Rev. 3
N° 1
12
This document may not be reproduced in partial form without the authorization of TAU STERIL.
8 Wiring Diagram
LEGEND
1.
Line Filter
2.
Fuse 5x20 extrafast 4 A.
3.
Fuse 5x20 semi-delayed 3.15 A.
4.
Fuse 5x20 semi-delayed 1.6 A.
5.
Transformer
6.
P.T.100 internal probe (temperature steam)
7.
Probe P.T.100 external (temperature resistance)
8.
pressure
9.
Attack printer connector
10.
Attack display connector
11.
Floating maximum discharge
12.
Floating minimum load
13.
Floating maximum load
14.
Floating common load connection
15.
Micro-switch door
16.
Electric valve N.A. (Normally open) 2 Way 24V DC.
17.
Output optional 24V. cc.
18.
Electric valve N.C. 2 Way 24V DC.
19.
Electric valve N.C. 2 Way 24V DC.
20.
230 V power supply (feeding the vacuum pump and the
electric valve nc 2 Vie)
21.
Bimetallic security
22.
Power supply 230 V 50 Hz.
23.
Resistance 230V 1250 Watts.
24.
switch
25.
Fuse 6.3x32 quick 10 A.
26.
Fuse 6.3x32 quick 10 A.
27.
Capacitor 0.1 mF 250 VAC class x2
28.
Cable clamp Power
29.
Mass ground
30.
Mass ground
31.
Electric valve N.C. 2 Way 24V DC.
32.
230 V power supply (feeds the water pump)
9 Control Panel
- LEGEND
1 = Pressure Display
2 = temperature display
3 = + - Temperature
4 = Display time
5 = + - Time
6 = Selection Program
7 = START / STOP
8 = Led Alarms
9 = Fuses
10 = Switch General
11 = Led Operation
12 = Led Program