TAU CLAVE 3000 Fractionated Vacuum 

Data: 30.03.15 

Pag. 5 di 14 

Rev. 3 

N° 1 

 

 

This document may not be reproduced in partial form without the authorization of TAU STERIL. 

 

 

 

 

 

 

3

 

Installation 

 
Carefully remove the autoclave from its packing and make sure there are no signs of damage. Make sure earthing is correctly 
made. Put the autoclave on a flat and horizontal surface, to facilitate aeration. Always respect a distance of at least 10 cm from 
any wall or other equipment, to allow suitable ventilation. 
Do not put objects or other items in the vicinity of the door, to prevent them from hindering opening of same. 
To check that the tension of suitable feeding on the back label corresponds to that in hand in the place of installation 
(maximum variation of the tension of net: ± 10%). 
To insert the thorn in the attack of the electric net. To press the general interrupter bringing him/it in position “I” and to verify 
the lighting of the same. 
In case of missed observance of these dispositions TAU STERIL declines every liability

 
 
 
 
 
 

                                         Fig. 2 

 
 
 
 
 
 

 
To give an inclination to the machine, it comes standard with two shims under the support legs front: it is important not to 
touch the shims. On the rear side of the machine is located in a pipe for the steam to exit. 

IMPORTANT: The tube must be positioned as shown in Figure 2. 

 

 
 
 
 
 
 
 
 
 
                                        

   Fig. 3                                                                              Fig. 4 

 

                   

 

-

 

Fill the bottle with 2.5 liters of demineralized water, then close the bottle with the cap (see Fig. 4). 

-

 

I

ntroduce the cannula orifice load autoclave. Unscrew the tap until the water begins to flow. The STOP appears on the 

display time, followed by a beep, it indicates that the machine is fully charged. 

-

 

When you turn on the red LED Lew level H2O, followed by the code A 01 (shown on the display time), it means the lack 
of water in the tank load.

 

-

 

Fill the bottle with 1.5 l of demineralized water.

 

-

 

Introduce the cannula orifice load the autoclave and unscrew the cap. The STOP appears on the display, which give you 
the completion of the load (about 1.5 l of the alarm activation A 01).

  

-

 

Tank drain recovery: Occurs alarm A 02 on the display of the time; It means that the machine must be emptied. 
Without this procedure, the same remains in the alarm condition, and therefore unusable. It is advisable for a 
better operation of the autoclave, the discharge of frequent recovery tank (dirty water).

 

-

 

Caution: It is recommended before using the autoclave with load (first load water happened) of looping 
"Unwrapped" (empty chamber without load) in order to sterilize the chamber.

 

 
 

Содержание Tau Clave 3000

Страница 1: ... TAU CLAVE 3000 Fractionated Vacuum Data 30 03 15 Pag 1 di 14 Rev 3 N 1 1 This document may not be reproduced in partial form without the authorization of TAU STERIL Instructions for Use Tau Clave 3000 Fractionated Vacuum 0426 DIRECTIVE 93 42 EEC and S M I MEDICAL DEVICES ...

Страница 2: ...ning and closure 4 3 Installation 5 4 Characteristics 6 4 1 Technical data 6 4 2 Environmental working conditions 6 4 3 Recommended times and characteristics of programmes 6 5 Operating modes 7 5 1 General information 7 5 2 Visualizations 7 5 3 Alarms visualization modes 8 5 3 1 Alarms and corresponding summary codes 8 5 4 Operating instructions 9 5 5 Description of Tau Clave 3000 Fr Vacuum cycle ...

Страница 3: ... CEI EN ISO 14971 The autoclave Tau 3000 Fractionated Vacuum complies with RoHS 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment 2 2 Expected use of materials to sterilize The equipment is used to sterilize surgical instruments in the medical paramedical sector The materials ...

Страница 4: ...O Instructions for Use TAU CLAVE 3000 Fractionated Vacuum Data 30 03 15 Pag 4 di 14 Rev 3 N 1 4 This document may not be reproduced in partial form without the authorization of TAU STERIL 2 4 Door opening and closure OPENING MOVEMENT CLOSING MOVEMENT ...

Страница 5: ... under the support legs front it is important not to touch the shims On the rear side of the machine is located in a pipe for the steam to exit IMPORTANT The tube must be positioned as shown in Figure 2 Fig 3 Fig 4 Fill the bottle with 2 5 liters of demineralized water then close the bottle with the cap see Fig 4 Introduce the cannula orifice load autoclave Unscrew the tap until the water begins t...

Страница 6: ... with phases of ventilation To select the cycle BOWIE DICK TEST as follows Place the program cycle HOLLOW before viewing 5 sec Of the cycle has finished press the button decreases the time Automatically displays the cycle 134 3 5 of BOWIE DICK TEST PROGRAM FRACTIONATED VACUUM STERILIZATION DRYING AND VENTILATION VACUUM FINAL UNWRAPPED Instruments unwrapped YES 134 4 Minutes 2 11 Bar 15 Minutes YES...

Страница 7: ...se the time in the program Special 6 INCREASE TIME to increase the time in the program Special 5 2 Views 1 PRESSURE It indicates the pressure in bar 2 TEMPERATURE Shows the temperature of steam in the chamber or by pressing the right button TEMP shows the temperature of the heating element Also pressing the left button TEMP you can see the selected temperature cycle 3 TIME Indicating the steriliza...

Страница 8: ...s alarm occurs during the cycle the latter is interrupted and the pressure discharged 5 3 1 Alarms and codes Summary CODE ALARM INDICATIONS A00 Door open occurs if the start of the cycle the door is open Disappears by pressing any key or by closing the door A01 Low water occurs if at the start of the cycle the water level in the cargo tank is insufficient Disappears by pressing any key or by reset...

Страница 9: ...pecial the contents of the display temperature will flash to indicate that the sterilization temperature can be chosen between 105 134 with the temperature The same applies to the time with the time These variations can be carried out simultaneously in both the time for which the temperature 7 Close the door after you have done so 8 Press Start starts the selected cycle 9 Keep area within the open...

Страница 10: ...sufficient ventilation Avoid placing the autoclave near sources of steam or splashing with water that may damage the internal electronics Do not install in places with poor ventilation Do not place it near heat sources 7 Maintenance and Cleaning Each type of maintenance must be carried out with power wire off and the machine completely cold WARNING Do not modify the equipment without the authoriza...

Страница 11: ...e at the side of the probe inside the sterilization chamber see fig 5 Fig 5 Dismantling With the cold machine turn from left to right the steel tube with the help of pliers if necessary paying attention to the probe closer Remove the piece from the chamber unscrew the filter clean it with vinegar or if too dirty replace it with a new one If you want you can avoid disassembling the tube with filter...

Страница 12: ...maximum load 14 Floating common load connection 15 Micro switch door 16 Electric valve N A Normally open 2 Way 24V DC 17 Output optional 24V cc 18 Electric valve N C 2 Way 24V DC 19 Electric valve N C 2 Way 24V DC 20 230 V power supply feeding the vacuum pump and the electric valve nc 2 Vie 21 Bimetallic security 22 Power supply 230 V 50 Hz 23 Resistance 230V 1250 Watts 24 switch 25 Fuse 6 3x32 qu...

Страница 13: ...and electrical equipment or put them out of order It s forbidden the operator to carry out repairs of any kind During cleaning and or maintenance it is absolutely essential to ensure that the equipment is disconnected from the power supply and it is completely cooled In the case of sterilization of surgical instruments eg scalpel or pointed instruments we recommend the use of safety shoes Periodic...

Страница 14: ...Rev 3 N 1 14 This document may not be reproduced in partial form without the authorization of TAU STERIL 10 2 Explanation of Symbols WARNING READ THE INSTRUCTIONS WARNING DANGER BURNS CAUTION RISK OF ELECTRIC SHOCK WARNING USE GLOVES FOR HAND PROTECTION PRODUCT NOT COMPARABLE TO MUNICIPAL WASTE FOLLOW THE RULES FOR MUNICIPAL DISPOSAL ...

Отзывы: