TAU STERIL s.n.c
Via Gorizia, 6/a
Fino Mornasco ( CO )
Instructions for Use
TAU CLAVE 3000 Fractionated Vacuum
Data: 30.03.15
Pag. 5 di 14
Rev. 3
N° 1
5
This document may not be reproduced in partial form without the authorization of TAU STERIL.
3
Installation
Carefully remove the autoclave from its packing and make sure there are no signs of damage. Make sure earthing is correctly
made. Put the autoclave on a flat and horizontal surface, to facilitate aeration. Always respect a distance of at least 10 cm from
any wall or other equipment, to allow suitable ventilation.
Do not put objects or other items in the vicinity of the door, to prevent them from hindering opening of same.
To check that the tension of suitable feeding on the back label corresponds to that in hand in the place of installation
(maximum variation of the tension of net: ± 10%).
To insert the thorn in the attack of the electric net. To press the general interrupter bringing him/it in position “I” and to verify
the lighting of the same.
In case of missed observance of these dispositions TAU STERIL declines every liability
.
Fig. 2
To give an inclination to the machine, it comes standard with two shims under the support legs front: it is important not to
touch the shims. On the rear side of the machine is located in a pipe for the steam to exit.
IMPORTANT: The tube must be positioned as shown in Figure 2.
Fig. 3 Fig. 4
-
Fill the bottle with 2.5 liters of demineralized water, then close the bottle with the cap (see Fig. 4).
-
I
ntroduce the cannula orifice load autoclave. Unscrew the tap until the water begins to flow. The STOP appears on the
display time, followed by a beep, it indicates that the machine is fully charged.
-
When you turn on the red LED Lew level H2O, followed by the code A 01 (shown on the display time), it means the lack
of water in the tank load.
-
Fill the bottle with 1.5 l of demineralized water.
-
Introduce the cannula orifice load the autoclave and unscrew the cap. The STOP appears on the display, which give you
the completion of the load (about 1.5 l of the alarm activation A 01).
-
Tank drain recovery: Occurs alarm A 02 on the display of the time; It means that the machine must be emptied.
Without this procedure, the same remains in the alarm condition, and therefore unusable. It is advisable for a
better operation of the autoclave, the discharge of frequent recovery tank (dirty water).
-
Caution: It is recommended before using the autoclave with load (first load water happened) of looping
"Unwrapped" (empty chamber without load) in order to sterilize the chamber.