3.5 Characters Time
11.3.5
Function Codes
04 (0x04) Read Input Registers
11.3.5.1
Data format
Tan Delta returns all real (non-integer) values as 16 bit signed integers with the value multiplied
by 100 (decimal).
For example, a temperature reading of 34.14 degrees C would be returned as 3414 decimal (0D56
hex).
Negative values follow the usual signed format, so -12.34 would be returned as -1234 decimal
(Fb2E hex).
11.3.5.2
Query
This reads the contents of input registers in the slave. The query message specifies the starting
register and number of registers to be read. Information needed to be read from the probe is
kept in registers:
Parameters
Registers
Decimal
Registers
Hex
Access
Oil Temperature *100
0
00
RO
Ambient Temperature * 100
1
01
RO
Oil Condition * 100
2
02
RO
Cal Zero * 100
3
03
RO
Oil Temperature deg F * 100
4
04
RO
Ambient Temperature deg F * 100
5
05
RO
Oil Condition TDN * 10
6
06
RO
Alarm State
7
07
RO
Encrypted RUL
8
08
RO
End of Life Value
10
0A
R/W
Instrument/Node Address
11
0B
R/W
Serial Type
(Modbus/CANbus/J1939)
12
0C
RO
Maximum Ambient Temp * 100
13
0D
RO
Serial No.
14
0E
RO
Hardware Version Number
15
0F
RO
Software Version Number * 100
16
10
RO
Serial No. High Digit
17
11
RO
Oil Serial No. 0 (Low)
18
12
RO
Oil Serial No. 1
19
13
RO
Oil Serial No. 2
20
14
RO
Oil Serial No. 3 (High)
21
15
RO
Oil Serial No. 4 (RESERVED)
22
16
RO
Oil Serial No. 5 (RESERVED)
23
17
RO
Oil Serial No. 6 (RESERVED)
24
18
RO
Содержание OQSExISG2
Страница 1: ...UG OQSxG2 23 0025 OIL QUALITY SENSOR ASSOCIATED ACCESSORIES USER GUIDE...
Страница 2: ......
Страница 5: ...Page 3 Page Intentionally Blank...
Страница 37: ...Page 35 Figure 19 Connection to other Tan Delta and 3rd Party Products Interfaces...
Страница 38: ...Page 36 Figure 20 ExOQSxG2 Ex Sensor Installation Drawing Rev 2 STAMPED...
Страница 42: ...Page 40 8 Display Express...
Страница 97: ...Page 95 P4 CANL RS485A P5 0V P6 CANH RS485B...