Edit (Edit/Delete/Copy/Import/Export Program)

4. On Screen Display and How to Use

Tab a protocol first, and then choose a function to proceed

Edit can be used to adjust the setting of existing 
protocols, which contains well volume, preheat temp, 
spin direction, extraction steps, time of mixing and collect.

To add a new program, you need to use Copy to 
duplicate a program first, and then use Edit to write the 
protocol.

Delete a program no longer used to save the memory 
space. Maelstrom 4810 can store up to 100 programs and 
each program can support totally 70 steps.

User could import the program into the instrument by 
USB.

Export function allows user to output the program by USB.

Edit function enables you to 
customize the extraction protocol.

Edit

ENG

LI

SH

Содержание Maelstrom 4810

Страница 1: ...Maelstrom 4810 User Manual TAN Bead Nucleic Acid Extractor Non Sterile 039 M1501 X01Rev V1 0_2021 06 11 Compatible with power 100V 240V...

Страница 2: ...ent Overview 11 3 Installation Get Started 15 4 On screen Display How to Use 16 5 Technical Support 36 6 Cleaning and Maintenance 37 7 Disposal 37 8 Patents 38 9 About Manufacturer 38 10 Appendix 39 2...

Страница 3: ...ty and environmental protection standards for products sold within the European Economic Area EEA In Vitro Diagnostic Medical Device Consult Instructions for Use Authorized Representative in the Europ...

Страница 4: ...the power cord Before working on maintenance must disconnect the power plug from the outlet in advance Do not pour any liquid on the instrument Do not place any containers with liquid on the instrumen...

Страница 5: ...so may cause malfunctions Repairs to the instrument must be performed by agencies authorized by Taiwan Advanced Nanotech Inc Only original spare parts from Taiwan Advanced Nanotech Inc can be used on...

Страница 6: ...etic rods to collect and transfer magnetic beads and spin tips to mix suspension Rotating the spin tips will mix suspension very well Extended magnetic rods can collect magnetic beads efficiently Puri...

Страница 7: ...culation Place Maelstrom 4810 on the table capable of bearing at least the weight of Maelstrom 4810 60 kg Do not use Maelstrom 4810 in a location where there is large temperature variability or high h...

Страница 8: ...should adhere to their institutional guidelines for the handling and disposal of all infectious substances used with this device It is important to clean the device after every use If samples or reag...

Страница 9: ...for electrical equipment for measurement control and laboratory use Part 2 101 Particular requirements for in vitro diagnostic IVD medical equipment EMC Requirements The medical device has passed the...

Страница 10: ...ccessories accompanying with instrument may vary from region to region Equipment list 10 1 Introduction Power cord x1 Remark Power cord varies from region to region Tips combo pack x1 Magnetic separat...

Страница 11: ...0 L x 430 W x 470 H mm Power rating AC100 AC240V 50 60Hz 5 2 5A Fuse 250V 5A Max Throughput 48 samples per run Process volume 50 l 1600 l Spin speed up to 3000 rpm Heater 12 independent heating blocks...

Страница 12: ...12 2 Instrument Overview ENGLISH HEPA Control panel Operation area Front and side view Power switch Power inlet Fuse...

Страница 13: ...trument Overview Control Panel To turn off the alarm Emergency to stop machine Please note that only TANBead technicians and authorized technicians or service engineers shall engage in program upgrade...

Страница 14: ...ox and must pay attention to the safety Step 1 Take out the instrument from the box 2 or more persons to handle is highly advised Step 2 Remove the cushions around the instrument Step 3 Place the inst...

Страница 15: ...code You can create new user code on the setting page User can only view their own and default code in code list Tab the blank space to enter the user code User ID New User ID Tab the blue arrow icon...

Страница 16: ...used on main page To mount and eject spin tips To execute the extraction program To export extraction records To use UV light to clean the instrument To add delete edit import export extraction progra...

Страница 17: ...tips will be ejected on the assigned lane which show on the Please be noticed that the lane you choose to eject the tips should be EMPTY Mount Eject assigned lane The tips in these lanes will be mount...

Страница 18: ...ing could be detected in Maelstrom 4810 When the alarm appear please check the tip status And remember 1 The door should be closed 2 The plate autotube should be placed in the right position Incomplet...

Страница 19: ...he program Or you could type the key words in the search slot to save the searching time Run function is designed to execute nucleic acid extraction program stored in the instrument Run 2 1 Kit and Ti...

Страница 20: ...characters input Once completed tab Next to continue The circle will turn red after the sample ID input successfully The input order will automatically turn to next number e g A1 will turn to A2 Row...

Страница 21: ...pre heating and plate type selection The plate type should be chosen in this page The type you chose will affect the auto eject tips or not The pre heating temp The status is fixed on 40 If you need...

Страница 22: ...is important It helps you clearly know the status of nucleic acid extraction From lane 1 to lane 12 name of buffer varies from kits to kits No of Well Name of Buffer Currently running well Currently b...

Страница 23: ...er you tab Pause or Stop button the instrument will not act immediately until the collection process comes to the end Vaporing EtOH Before entering Elute process evaporating EtOH is a MUST Poor evapor...

Страница 24: ...ecords and the identity that performed the experiment 24 Report Export extraction records 4 On Screen Display and How to Use The record could be chosen by tabbing the square In front of date The chose...

Страница 25: ...1 Set lighting time by clicking UV duration field Step 2 Click bulb icon to begin UV lighting Before turn on UV please must close the door of operation area The recommended lighting time is at least 3...

Страница 26: ...time of mixing and collect To add a new program you need to use Copy to duplicate a program first and then use Edit to write the protocol Delete a program no longer used to save the memory space Maels...

Страница 27: ...the mixing process For Forward spin clockwise Rev Reverse spin spin direction switch regularly For U D Forward spin with up and down move Rev U D Reverse spin with up and down move Name Name of buffe...

Страница 28: ...mixing range from 0 0 to 60 0 mins Speed Speed of mixing range from 500 to 3500 rpm Collect Time of collecting beads range from 0 0 to 60 0 mins Vapor Time of evaporate EtOH range from 0 0 to 60 0 min...

Страница 29: ...otocol Confirm the number to add the program The keyboard will appear after tab on destination number Once complete click Yes to add the program Please find a number to add a program Then tab a progra...

Страница 30: ...delete a program tab the program no longer need and then click Delete icon to execute Click Yes to confirm the deletion 2 1 Tab the unwanted program and tab the Delete icon 30 4 On Screen Display and...

Страница 31: ...language display including English Traditional Chinese Simplified Chinese Korean and Spanish Time When a new instrument arrives in we strongly suggest user setting right time for the instrument Barcod...

Страница 32: ...same as 333 2 1 Tab User icon to access Setting User management User can log in by default code 333 or create their own code in this function The clearance will be same as 333 and each user could only...

Страница 33: ...isplay language 2 1 Tab Language icon to access Setting Language Maelstrom 4810 supports multi language operating interface including English Traditional Chinese Simplified Chinese Korean and Espa ol...

Страница 34: ...Setting Time Set correct time for the system according to the local time Click icon to set Date and Time for the system 34 4 On Screen Display and How to Use ENGLISH 2 2 1...

Страница 35: ...h USB type infrared barcode scanner Click the icon to ON OFF of barcode scanner If barcode scanner is set OFF even USB type barcode scanner is connected to Maelstrom 4810 scanned data will not be pres...

Страница 36: ...ur company for more help If there is a need to contact us you can call our rep number at 886 3 3167568 or email us at service tanbead com for further assistance To efficiently solve the problems and a...

Страница 37: ...rods unprotected If this happens the magnetic rod needs to be cleaned immediately To clean the magnetic rods wipe with a soft cloth dampened with pure water Do not use alcohol solvent If the magnetic...

Страница 38: ...511 3908 9530 9 About Manufacturer Patent List USA US09616398B2 EU EP2937136 Canada CA2862946 Japan JP6151735B2 Korea KR101696517B1 China CN104971638B Taiwan TWI526245B WIPO WO2016127292 8 Patent Man...

Страница 39: ...e below to replace the fuses 1 Disconnect the power to the instrument 2 Unplug the power cord from the instrument 3 Still the instrument at least 1 min before fuse replacement Procedure of Fuse Replac...

Страница 40: ...6F No 188 Wenhe Rd Guishan Dist Taoyuan City 333 Taiwan R O C Tel 886 3 316 7568 Fax 886 3 317 3369 www tanbead com Taiwan Advanced Nanotech Inc...

Отзывы: