Tamson Instruments TV12 Скачать руководство пользователя страница 9

 

 

www.tamson.com 
[email protected]

 

 

van 't Hoffstraat 12 
2665 JL Bleiswijk, The Netherlands 
T. 31 (0) 10 522 43 73 

tv12-man Rev 2.0 UK 1017 

 

Tamson Instruments bv 

Page 9/29 

 

ISO 9001 : 2015 
NL/PRO 238239125 

 

VAT: NL 80 66 34 984 B01 

Bank account no.:  

NL28 INGB 0007 350 370 

NL95 RABO  0160100046 

Chamber of commerce 27 16 95 41 

 
 
 
 

 
 

4.6  Bath fluids 

We recommend the use of  oil with viscosity < 10 mm²/s 
 

Recommended bath fluids 
Range 

Ordering code 

Description 

Ambient to  
80

C/176

N.A. 

Clean tap water  
NOT DISTILLED  

80..150

C / 

176..302

00T0220 (20 litres) 

Tamson mineral oil 150. Transparent, 20 mm²/s @ 80°C/176°F; 3 
mm²/s @ 150°C/302°F; F.P. 200°C/392°F. 

20..150

C / 

68..302

08T0001 (gross 20 
Kg = 20 ltrs) 

Silicon oil 200-10, transparent. 10 mm²/s @ 25°C/77°F;10 mm²/s 
@ 150°C/301°F. F.P. 211°C/411°F. Lifetime >2 years. 

 
 
 
 
 

The oil used has a limited lifetime.  

 
 
The  type,  brand  and  operating  temperature  mainly 
determines lifetime. Spilling of sample may also reduce 
lifetime,  in  some  cases  can  start  chemical  reactions. 
Silicon  oil  has  the  tendency  to  form  gel,  for  this  reason 
silicon oil has to be replaced as soon as visible changes 
are noticed like string forming. Within a few hours silicon 
oil can transform itself into solid gel, which is very difficult 
to remove. When not totally cleaned, very small pieces of 
left over gel will catalyze new oil to form gel! 
 
 

Do  not  use  distilled  water.  This  will  cause  serious 
corrosion.

 

 

 

Only  use  water  as  a  bath  fluid  below  80°C/176°F. 
Working  for  a  longer  period  with  water  at 
temperatures  above  80°C  will  damage  the  stirrer 
bearings.  

 

 

 

Using the bath for long periods at low temperatures 
will  cause  condensation  on  the  glass  window 
resulting  in  complete  invisibility  of  material  placed 
inside the bath. 

 
 
 
 

 

CHECK  

EXPIRING DATE 

OIL 

AFTER  LONG  

PERIODS OF LOW 

TEMPERATURES 

VISIBILITY BECOMES 

LESS  

NO TEMPERATURE  

> 80ºC 

H2O 

Содержание TV12

Страница 1: ...swijk The Netherlands T 31 0 10 522 43 73 tv12 man Rev 2 0 UK 1017 Tamson Instruments bv Page 1 29 ISO 9001 2015 NL PRO 238239125 VAT NL 80 66 34 984 B01 Bank account no NL28 INGB 0007 350 370 NL95 RA...

Страница 2: ...UCTION 13 6 3 STIRRER 13 6 4 TEMPERATURE CONTROL AND SETTING 13 6 5 SAFETY SYSTEMS 14 6 6 ADJUSTING THE MECHANICAL OVER TEMPERATURE PROTECTION THERMOSTAT 14 6 7 FRONT PANEL 14 6 8 OVERVIEW MENU ITEMS...

Страница 3: ...quipment or its accessories caused by incorrect operation Important When performing service maintenance or moving the apparatus always disconnect the apparatus at the mains socket Proper skilled and t...

Страница 4: ...condensating Atmosphere Not suited for altitudes above 2000m or explosive corrosive environment Pollution cat 2 Conducting pollution must be prevented 4 INSTALLATION 4 1 Important Tamson Instruments...

Страница 5: ...38239125 VAT NL 80 66 34 984 B01 Bank account no NL28 INGB 0007 350 370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 holders etc Please see the packing list for details concerning total conten...

Страница 6: ...fluid is low enough at the operating temperature The viscosity must be below 10 mm s High viscosity will result in poor stability as well as poor uniformity of the bath The fluid flash point must be w...

Страница 7: ...evel Minimum level The heating element must be fully covered by the bath fluid The heating element will be damaged when not fully submerged in the bath fluid The practical minimum level is indicated b...

Страница 8: ...If the fluid level is too high it will leave the bath via the overflow outlet 14mm outher diameter pipe Prevent fluid from the overflow outlet entering the backside of the TV12 For this reason the ov...

Страница 9: ...ifetime The type brand and operating temperature mainly determines lifetime Spilling of sample may also reduce lifetime in some cases can start chemical reactions Silicon oil has the tendency to form...

Страница 10: ...hot Always use heat protective gloves Care must be taken when placing or re moving material from the bath Pay attention when removing hot fluid Use heat protective clothing and wear gloves and safety...

Страница 11: ...9125 VAT NL 80 66 34 984 B01 Bank account no NL28 INGB 0007 350 370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 Do not spill water in hot oil Never empty hot oil into a plastic container Heat...

Страница 12: ...below ambient it is necessary to pass a small amount of tap water through the cooling coil Generally the temperature of the cooling coil water should be 5 degrees or more below the desired operating...

Страница 13: ...rature distribution within the bath 6 4 Temperature control and setting The bath temperature is regulated using a Pt 100 temperature probe Class A connected to a microprocessor module The advanced ele...

Страница 14: ...example oil to be heated above flaming point which will cause fire or prevent evaporation of bath fluid due to high temperatures The thermostat will automatically reset when the bath temperature drops...

Страница 15: ...m High alarm PID parameter PID set 1 PID set 2 PID set 3 PID set 4 Each PID set offers settings for Proportional band value Pb 1 P where P is proportional value Integral value Differential value Backl...

Страница 16: ...isplayed value increases to a stable readout appears after a few seconds Ad 2 The controller calculates every second the amount of power which should be applied for stable control The value is display...

Страница 17: ...en acquired using following PID settings Pb 25 I 16 D 0 Under different settings its possible to achieve even better values by trimming the PID settings Place the bath spirit level The four supporting...

Страница 18: ...d to select time to tune The option precise has to be used to reach maximum temperature accuracy Options are Fast 60 seconds Medium 120 seconds Slow 180 seconds Precise 240 seconds Standard value Prec...

Страница 19: ...350 370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 On Off Standard value On Menu item Temp units C F Standard value C Menu item Baudrate 300 600 1200 2400 4800 9600 19200 38400 Standard val...

Страница 20: ...L95 RABO 0160100046 Chamber of commerce 27 16 95 41 8 1 PID settings Following shows influence on stability with different PID settings Temperature 75 C Proportional band Pb 25 Integrator 16 Different...

Страница 21: ...with the use of the Ziegler Nichols method described below With the proces at its normal proces value Set the integral time Ti and the derivative Time Td to off Check if the Lcb and Hcb are set to aut...

Страница 22: ...46 Chamber of commerce 27 16 95 41 8 3 Precision of control Temperature 40 C Proportional band Pb 25 Integrator 16 Differentiator 0 Min max 0 008 C Temperature 50 C Proportional band Pb 25 Integrator...

Страница 23: ...007 350 370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 Temperature settling times When the bath temperature is disturbed i e when viscometer holders are placed the electronic regulation will...

Страница 24: ...is distributed rather randomly More sophisticated systems have controlled flow by using a baffle plate This still does not prevent horizontal and vertical temperature fluctuations The special heating...

Страница 25: ...is not turning The motor fuse is activated Restart the motor by pressing the motor fuse This fuse is located at the backside of the top casing Also check viscosity of the bath fluid High viscosity wi...

Страница 26: ...h the cooling coil If the operating temperature is far above the ambient temperature it is most likely that the viscosity of the bath fluid used is too high The maximum viscosity of the bath fluid lie...

Страница 27: ...e 0 6 Amp 06T0512 06T0522 PCB Mains board with filter and relay 25T0351 Heater W 24T8581 Over temperature protection thermostat 24T8545 Mains switch 24T8546 Protective cover mains switch 28T4019 Front...

Страница 28: ...HS Directive 2011 65EU p b EN61010 2 010 Safety requirements for electrical equipment for measurement control and laboratory use Particular require ments for laboratory equipment for the heating of ma...

Страница 29: ...Conditions of use and suitability of the product for particular uses are beyond our control all risks of use of the product are therefore assumed by the user We expressly disclaim all warranties of ev...

Отзывы: