The front panel layout shows the turn-push button:
Next / increase:
Turn right
Previous / decrease:
Turn left
Select:
Press
6.8 Overview menu items
-
Set point
-
Offset (press: <-5.00 .. +5.00°C resolution 0.01°C)
-
Max Power (press: low 25, med, hi, max)
-
Boost heater (press on / off)
-
Time const (press: fast, medium slow, precise)
-
Stirrer
-
Low alarm
-
High alarm
-
PID parameter :
PID set 1,
PID set 2,
PID set 3,
PID set 4.
Each PID set offers settings for
- Proportional band value
(Pb=1/P where P is proportional value)
- Integral value
- Differential value
-
Backlight
-
Temp units
-
Baudrate
-
SP Offset
-
Restart
6.9 Safety thermostat
When the bath temperature becomes higher than the set
point temperature of the safety thermostat (Indicated as
7, see page 14), the heating electronics and controller are
switched-off. The red LED on the display will light. When
the temperature of the bath has been lowered with
approximately 10°C, the thermostat re-sets itself
automatically, however to continue normal operation the
bath has to be switched off and on again.
Use well grounded mains. Before plugging the TV12 into the mains socket, make sure the voltage of the bath
corresponds to the local voltage.