![Tamson Instruments TV12 Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/tamson-instruments/tv12/tv12_user-manual_3451932012.webp)
www.tamson.com
[email protected]
van 't Hoffstraat 12
2665 JL Bleiswijk, The Netherlands
T. 31 (0) 10 522 43 73
tv12-man Rev 2.0 UK 1017
Tamson Instruments bv
Page 12/29
ISO 9001 : 2015
NL/PRO 238239125
VAT: NL 80 66 34 984 B01
Bank account no.:
NL28 INGB 0007 350 370
NL95 RABO 0160100046
Chamber of commerce 27 16 95 41
5
Cooling
The bath is provided with an integral cooling coil.
Because of the friction-heat generated by the stirrer
mechanism the bath will heat-up slowly. The lowest
operating temperature at which the bath can be controlled
depends on the fluid used and the ambient temperature.
When operating at temperatures around or below
ambient it is necessary to pass a small amount of tap
water through the cooling coil. Generally the temperature
of the cooling coil water should be 5 degrees (or more)
below the desired operating temperature of the bath. To
operate around ambient temperature a flow of 1 to 2l/min.
is sufficient.
Instead of using tap water a Tamson cooling circulator
can be connected to the TV12 to provide cooling. The
TLC10-3 is our smallest circulator and well suited for this
purpose.
5.1 Connecting
Before plugging TV12 into the mains socket, make sure
the voltage of the bath corresponds to the local voltage.
Use a mains supply that is well earthed, clean of
interference and suitable for the acquired electrical
load of the bath.
CONNECTORS FOR
COOLING
(OUTSIDE DIAMETER 10MM)
CHECK VOLTAGE
USE WELL EARTHED MAINS
TLC10-3
COOLING CIRCULATOR