6

 

Note: 

It is important that during the INITIALISING phase the mattress connector is not 

disconnected from the power unit.  If this is done, the power unit must be switched off, 
MUTE button pressed when the sounder is heard, the mattress connector re-engaged, 
and the power unit restarted. If UNCALIBRATED is displayed when switched on, the 
system will continue to operate but should be recalibrated as soon as possible.

9.  When a bottom sheet is added to the mattress, ensure that it is left loose to allow the 

mattress surface to conform with the patient's body as much as possible.

10.  Allow the mattress to inflate before positioning the patient on the mattress.

11.  Once inflated the system will automatically default to ACTIVE mode. CONTINUOUS 

LOW PRESSURE therapy can be selected via the THERAPY MODE button (please see 
THERAPY MODE on page 7). The comfort setting can be adjusted using the up and down 
arrow buttons (see COMFORT CONTROL on page 7).

NB. The power unit will automatically lock 2 minutes after last button operation 
when running to prevent inadvertent operation of button functions (except MUTE), as 
indicated by     on the display screen. Press and hold the MUTE/UNLOCK button until 
power unit beeps if further button operation is needed (i.e. change of therapy mode or 
comfort setting).

12.  Place the carry bag and user manual in a safe place for future use.

INSTALLING CUSHIONS

The QUATTRO

®

 PLUS, QUATTRO

®

 OVERLAY AND PULSAIR

®

 CHOICE power units can all be used 

to operate their respective alternating air pressure cushions. Place the cushion on a chair, 
ensuring that it is placed the correct way up with the BACK labelling (if applicable) facing the 
back of the chair.  Secure cushion to chair using adjustable straps, where available.

1.  The PULSAIR

®

 CHOICE cushion is connected to the PULSAIR

®

 CHOICE power unit exactly 

as described in points 5. to 8. above and will take approx. 30 seconds to inflate.

2.  The B.A.S.E.

®

 SEQUENTIAL and B.A.S.E.

®

 RECLINER MAT cushions can be used with the 

QUATTRO

®

 PLUS and QUATTRO

®

 OVERLAY  mattress systems, either operated with the 

mattress or, using a cushion adaptor, directly with the power unit, as follows:-

 

a) To use the cushion directly with the power 

unit, connect the cushion adaptor to the 
power unit, rotating the adaptor clockwise 
until the green indicator fills the indicator 
hole on the power unit connector. Connect 
the air supply hose from the cushion onto 
the adaptor, matching up the alignment 
markings (Fig. 3). Ensure that this has been

Fig. 3            

Содержание B.A.S.E RECLINER MAT

Страница 1: ...Pressure Area Care Mattresses and Cushions QUATTRO ACUTE Systems QUATTRO PLUS Systems QUATTRO OVERLAY Systems PULSAIR CHOICE Systems POLYFLOAT SUPREMA POLYFLOAT DORMIRA B A S E Seating Systems reduci...

Страница 2: ...the user to the possibility of a problem with the system associated with its use or misuse Operating Instructions IP Ingress Protection 2 Protection against fingers or other object not greater than 80...

Страница 3: ...n a cot size version for paediatric use l QUATTRO PLUS mattress replacement system for patients at high risk of pressure ulcers l QUATTRO OVERLAY mattress overlay system for patients at high risk of p...

Страница 4: ...sing or lowering of bed or bed rails or any other moving object l The power unit must only be used with a suitable approved cord and plug set as supplied by Talley l The system is not used in the pres...

Страница 5: ...injury l In order for alternating air pressure mattress cushion range to operate effectively please avoid placing objects on the surface that may obstruct the movement of air between the cells For th...

Страница 6: ...d mattress and causing possible injury to the patient A strap extension kit is available for use with divan beds part number 97 50 10 153 4 Suspend the power unit from the foot board of the bed having...

Страница 7: ...screen Press and hold the MUTE UNLOCK button until power unit beeps if further button operation is needed i e change of therapy mode or comfort setting 12 Place the carry bag and user manual in a safe...

Страница 8: ...The mattress and cushion will operate simultaneously allowing the simple transfer of the patient between the bed and chair NB When the patient is in bed return the comfort setting to the mattress set...

Страница 9: ...The power unit will lock again 2 minutes after the last button operation NB After power failure switching the power off pressing MUTE cancels the system s previous settings When power returns the def...

Страница 10: ...ain inflated so supporting the patient All alternating air pressure mattresses can be used on profiling bed frames slatted frames in filled frames and divans Back rests or pillows for support should b...

Страница 11: ...insed with clean water to remove any residues This will help prevent any possible long term compatibility issues associated with disinfectant residues l Alternatively disinfection may be achieved by l...

Страница 12: ...iversally used on all hard and soft surfaces without any detrimental effect TECcare CONTROL is CE marked for cleaning medical equipment SERVICING Once the initial guarantee period expires Talley recom...

Страница 13: ...en All systems have a fault log that records the last 5 faults via the DATA display mode If problems reoccur contact Talley AC FAIL fault indicates a mains power failure a sounder will be heard if pow...

Страница 14: ...by pressing MUTE button EMI FAULT Caused by external RF fields Fault will clear when interference ceases Check air cells and tubing are intact FAULT Reset power unit by pressing MUTE button MATTRESS...

Страница 15: ...gap of 2 5cm either side of the mattress should not be exceeded when side rails are deployed CUSHIONS Place the POLYFLOAT cushion onto a chair ensuring cushion is placed the correct way up N B It shou...

Страница 16: ...s will help prevent any possible long term compatibility issues associated with disinfectant residues l Alternatively disinfection may be achieved by laundering at temperatures not exceeding 650 C for...

Страница 17: ...Time QUATTRO ACUTE variable continuous QUATTRO PLUS 16 minutes continuous QUATTRO OVERLAY 16 minutes continuous PULSAIR CHOICE 16 minutes mattress replacement 8 minutes mattress overlay cushion conti...

Страница 18: ...30mm x 430mm x 70mm UK standard size B A S E RECLINER MAT cushion 1080mm x 430mm x 70mm PULSAIR CHOICE cushion 430mm x 430mm x 40mm Weight B A S E SEQUENTIAL cushion 1 9kg n B A S E RECLINER MAT cushi...

Страница 19: ...ering equipment repaired or replaced l Reorient or relocate the interfering equipment l Increase the separation between the equipment and the possible source of the interference l Connect the equipmen...

Страница 20: ...separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interfer...

Отзывы: