Talley B.A.S.E RECLINER MAT Скачать руководство пользователя страница 3

Contents

 

         

 

Page

EXPLANATION OF LABELS SYMBOLS AND STATEMENTS 

1

INTRODUCTION 2
CAUTIONS AND WARNINGS INFORMATION 

3

ALTERNATING AIR PRESSURE MATTRESSES & CUSHIONS

INSTALLATION AND USER GUIDELINES 

5

CARE AND MAINTENANCE 

10

FAULT FINDING 

12

FOAM MATTRESSES & CUSHIONS

INSTALLATION AND USER GUIDELINES 

14

CARE AND MAINTENANCE 

14

SPECIFICATIONS 16
EMI/EMC STATEMENT AND MANUFACTURER’S DECLARATION 18

Introduction

Thank you for choosing to use a Talley pressure area care product, effective for the prevention 
and management of pressure ulcers.

The range of alternating air pressure products is as follows:-

QUATTRO

®

 ACUTE: mattress replacement system for patients at very high risk of pressure 

ulcers (also available in a cot size version for paediatric use)

QUATTRO

®

 PLUS: mattress replacement system for patients at high risk of pressure ulcers

QUATTRO

®

 OVERLAY: mattress overlay system for patients at high risk of pressure ulcers

PULSAIR

®

 CHOICE mattress replacement: for patients at high risk of pressure ulcers

PULSAIR

®

 CHOICE mattress overlay: for patients at medium risk of pressure ulcers 

PULSAIR

®

 CHOICE cushion: for patients at medium risk of pressure ulcers

B.A.S.E.

®

 SEQUENTIAL: seat cushion for patients at very high risk of pressure ulcers

B.A.S.E.

®

 RECLINER MAT

: seat and back cushion for patients at very high risk of pressure 

ulcers 

The range of static foam products is as follows:-

POLYFLOAT SUPREMA

: mattress replacement for patients at risk of pressure ulcers

POLYFLOAT DORMIRA

: mattress replacement for patients at risk of pressure ulcers

POLYFLOAT

 cushion: for patients at risk of pressure ulcers 

2

Содержание B.A.S.E RECLINER MAT

Страница 1: ...Pressure Area Care Mattresses and Cushions QUATTRO ACUTE Systems QUATTRO PLUS Systems QUATTRO OVERLAY Systems PULSAIR CHOICE Systems POLYFLOAT SUPREMA POLYFLOAT DORMIRA B A S E Seating Systems reduci...

Страница 2: ...the user to the possibility of a problem with the system associated with its use or misuse Operating Instructions IP Ingress Protection 2 Protection against fingers or other object not greater than 80...

Страница 3: ...n a cot size version for paediatric use l QUATTRO PLUS mattress replacement system for patients at high risk of pressure ulcers l QUATTRO OVERLAY mattress overlay system for patients at high risk of p...

Страница 4: ...sing or lowering of bed or bed rails or any other moving object l The power unit must only be used with a suitable approved cord and plug set as supplied by Talley l The system is not used in the pres...

Страница 5: ...injury l In order for alternating air pressure mattress cushion range to operate effectively please avoid placing objects on the surface that may obstruct the movement of air between the cells For th...

Страница 6: ...d mattress and causing possible injury to the patient A strap extension kit is available for use with divan beds part number 97 50 10 153 4 Suspend the power unit from the foot board of the bed having...

Страница 7: ...screen Press and hold the MUTE UNLOCK button until power unit beeps if further button operation is needed i e change of therapy mode or comfort setting 12 Place the carry bag and user manual in a safe...

Страница 8: ...The mattress and cushion will operate simultaneously allowing the simple transfer of the patient between the bed and chair NB When the patient is in bed return the comfort setting to the mattress set...

Страница 9: ...The power unit will lock again 2 minutes after the last button operation NB After power failure switching the power off pressing MUTE cancels the system s previous settings When power returns the def...

Страница 10: ...ain inflated so supporting the patient All alternating air pressure mattresses can be used on profiling bed frames slatted frames in filled frames and divans Back rests or pillows for support should b...

Страница 11: ...insed with clean water to remove any residues This will help prevent any possible long term compatibility issues associated with disinfectant residues l Alternatively disinfection may be achieved by l...

Страница 12: ...iversally used on all hard and soft surfaces without any detrimental effect TECcare CONTROL is CE marked for cleaning medical equipment SERVICING Once the initial guarantee period expires Talley recom...

Страница 13: ...en All systems have a fault log that records the last 5 faults via the DATA display mode If problems reoccur contact Talley AC FAIL fault indicates a mains power failure a sounder will be heard if pow...

Страница 14: ...by pressing MUTE button EMI FAULT Caused by external RF fields Fault will clear when interference ceases Check air cells and tubing are intact FAULT Reset power unit by pressing MUTE button MATTRESS...

Страница 15: ...gap of 2 5cm either side of the mattress should not be exceeded when side rails are deployed CUSHIONS Place the POLYFLOAT cushion onto a chair ensuring cushion is placed the correct way up N B It shou...

Страница 16: ...s will help prevent any possible long term compatibility issues associated with disinfectant residues l Alternatively disinfection may be achieved by laundering at temperatures not exceeding 650 C for...

Страница 17: ...Time QUATTRO ACUTE variable continuous QUATTRO PLUS 16 minutes continuous QUATTRO OVERLAY 16 minutes continuous PULSAIR CHOICE 16 minutes mattress replacement 8 minutes mattress overlay cushion conti...

Страница 18: ...30mm x 430mm x 70mm UK standard size B A S E RECLINER MAT cushion 1080mm x 430mm x 70mm PULSAIR CHOICE cushion 430mm x 430mm x 40mm Weight B A S E SEQUENTIAL cushion 1 9kg n B A S E RECLINER MAT cushi...

Страница 19: ...ering equipment repaired or replaced l Reorient or relocate the interfering equipment l Increase the separation between the equipment and the possible source of the interference l Connect the equipmen...

Страница 20: ...separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interfer...