This medical device is compliant with:

IEC 60601.1 3rd edition Medical electrical equipment safety and essential performance

IEC 60601.1.11 Home healthcare environment

Guidance and Manufacturer’s Declaration: Electromagnetic Immunity (IEC 60601-1-2)

Immunity Test

IEC 60601 Test Level

Compliance Level

Electromagnetic Environment - Guidance

Electrostatic 

discharge (ESD) 

IEC 61000-4-2

± 6kV contact

± 8kV air

± 6kV contact

± 8kV air

Floors should be wood, concrete or ceramic tile. If floors 

are covered with synthetic material, the relative humidity 

should be at least 30%.

Electrical fast 

transient/burst IEC 

61000-4-4

± 2 kV For mains supply lines

± 1kV For input/output lines

± 2 kV For mains supply lines

± 1kV For input/output lines

Mains supply quality for the mains adapter should be 

that of a typical commercial and/or hospital environment.

Surge

IEC61000-4-5

± 1kV line(s) to line

± 1kV line(s) to line

Mains supply quality for the mains adapter should be 

that of a typical commercial and/or hospital environment.

Voltage dips, short 

interruptions and 

voltage variations 

on mains supply 

IEC 61000-4-11

<5%Ur (>95%Ur) for 0.5 cycle 

<5%Ur (>95%Ur) for 0.5 cycle 

Mains power quality should be that of a typical 

commercial and/or hospital environment. If the user of 

the QUATTRO

®

 requires continued operation during 

power mains interruption, it is recommended that the 

QUATTRO

®

 be powered from an uninterruptible power 

supply or battery.

40%Ur (60% dip in Ur) for 5 cycles

40%Ur (60% dip in Ur) for 5 cycles

70%Ur (30% dip in Ur) for 25 cycles 70%Ur (30% dip in Ur) for 25 cycles
>5%Ur (>95% dip in Ur) for 5 secs

>5%Ur (>95% dip in Ur) for 5 secs

Mains frequency 

(50/60Hz) magnetic 

field IEC61000-4-8

3 A/m

3 A/m

Mains frequency magnetic fields should be at levels 

characteristic of a typical location in a typical commercial 

and/or hospital environment.

Note: Ur is the A.C. mains voltage prior to application of the test level.

Guidance and Manufacturer’s Declaration: Electromagnetic Immunity (IEC 60601-1-2)

Immunity 

Test

IEC 60601 

Test Level

Compliance 

Level

Electromagnetic Environment - Guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V rms

150 kHz ~ 80 MHz

3 V/m

80 MHz ~

2.5 GHz

3 V rms

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part of the 

QUATTRO

®

 including cables, than the recommended separation distance calculated from the 

equation appropriate to the frequency of the transmitter.

Recommend separation distance

d = 1.2 √P

d = 1.2 √P 80 MHz to 800 MHz

d = 2.3 √P 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the 

transmitter manufacturer and d is the recommended separation

distance in meters (m).

Field strengths from fixed RF transmitters as determined by an electromagnetic site

survey,

a

 should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

Note1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and 

people.

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radio, AM and FM radio broadcast, 

and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electro

-

magnetic site survey should be considered. If the measured field strength in the location in which the 

QUATTRO

®

 is used exceeds the applicable RF 

compliance level above, the QUATTRO

®

 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be 

necessary, such as reorienting or relocating the QUATTRO

®

.

b

 Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.

Recommended Separation Distance Between Portable and Mobile RF Communications Equipment and the QUATTRO

®

The QUATTRO

®

 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the QUATTRO

®

 can 

help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) 

and the QUATTRO

®

 as recommended below, according to the maximum output power of the communications equipment.

Output Power of Transmitter in

Watts (W)

Separation distance according to frequency of transmitter in Meters (m)

150 kHz to 80 MHz d = 1.2 √P

80 MHz to 800 MHz d = 1.2 √P

800 MHz to 2.5GHz d = 2.3 √P

0.01

0.01

0.01

0.01

0.12

0.12

0.12

0.12

0.12

0.12

0.12

0.12

0.23

0.23

0.23

0.23

0.1

0.1

0.1

0.1

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in Meters (m) can be estimated using the 

equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to the 

transmitter manufacturer. Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies. Note: These guidelines may not 

apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

USER MANUAL PART NUMBER 50-02-07-200/13

02/2017

Talley Group Limited

Premier Way, Abbey Park Industrial Estate, Romsey, Hampshire, SO51 9DQ England

TEL

+

44(0)1794 503500   

FAX

:  

+44(0)1794 503555

E-MAIL

:  

[email protected]

www.talleygroup.com

Содержание B.A.S.E RECLINER MAT

Страница 1: ...Pressure Area Care Mattresses and Cushions QUATTRO ACUTE Systems QUATTRO PLUS Systems QUATTRO OVERLAY Systems PULSAIR CHOICE Systems POLYFLOAT SUPREMA POLYFLOAT DORMIRA B A S E Seating Systems reduci...

Страница 2: ...the user to the possibility of a problem with the system associated with its use or misuse Operating Instructions IP Ingress Protection 2 Protection against fingers or other object not greater than 80...

Страница 3: ...n a cot size version for paediatric use l QUATTRO PLUS mattress replacement system for patients at high risk of pressure ulcers l QUATTRO OVERLAY mattress overlay system for patients at high risk of p...

Страница 4: ...sing or lowering of bed or bed rails or any other moving object l The power unit must only be used with a suitable approved cord and plug set as supplied by Talley l The system is not used in the pres...

Страница 5: ...injury l In order for alternating air pressure mattress cushion range to operate effectively please avoid placing objects on the surface that may obstruct the movement of air between the cells For th...

Страница 6: ...d mattress and causing possible injury to the patient A strap extension kit is available for use with divan beds part number 97 50 10 153 4 Suspend the power unit from the foot board of the bed having...

Страница 7: ...screen Press and hold the MUTE UNLOCK button until power unit beeps if further button operation is needed i e change of therapy mode or comfort setting 12 Place the carry bag and user manual in a safe...

Страница 8: ...The mattress and cushion will operate simultaneously allowing the simple transfer of the patient between the bed and chair NB When the patient is in bed return the comfort setting to the mattress set...

Страница 9: ...The power unit will lock again 2 minutes after the last button operation NB After power failure switching the power off pressing MUTE cancels the system s previous settings When power returns the def...

Страница 10: ...ain inflated so supporting the patient All alternating air pressure mattresses can be used on profiling bed frames slatted frames in filled frames and divans Back rests or pillows for support should b...

Страница 11: ...insed with clean water to remove any residues This will help prevent any possible long term compatibility issues associated with disinfectant residues l Alternatively disinfection may be achieved by l...

Страница 12: ...iversally used on all hard and soft surfaces without any detrimental effect TECcare CONTROL is CE marked for cleaning medical equipment SERVICING Once the initial guarantee period expires Talley recom...

Страница 13: ...en All systems have a fault log that records the last 5 faults via the DATA display mode If problems reoccur contact Talley AC FAIL fault indicates a mains power failure a sounder will be heard if pow...

Страница 14: ...by pressing MUTE button EMI FAULT Caused by external RF fields Fault will clear when interference ceases Check air cells and tubing are intact FAULT Reset power unit by pressing MUTE button MATTRESS...

Страница 15: ...gap of 2 5cm either side of the mattress should not be exceeded when side rails are deployed CUSHIONS Place the POLYFLOAT cushion onto a chair ensuring cushion is placed the correct way up N B It shou...

Страница 16: ...s will help prevent any possible long term compatibility issues associated with disinfectant residues l Alternatively disinfection may be achieved by laundering at temperatures not exceeding 650 C for...

Страница 17: ...Time QUATTRO ACUTE variable continuous QUATTRO PLUS 16 minutes continuous QUATTRO OVERLAY 16 minutes continuous PULSAIR CHOICE 16 minutes mattress replacement 8 minutes mattress overlay cushion conti...

Страница 18: ...30mm x 430mm x 70mm UK standard size B A S E RECLINER MAT cushion 1080mm x 430mm x 70mm PULSAIR CHOICE cushion 430mm x 430mm x 40mm Weight B A S E SEQUENTIAL cushion 1 9kg n B A S E RECLINER MAT cushi...

Страница 19: ...ering equipment repaired or replaced l Reorient or relocate the interfering equipment l Increase the separation between the equipment and the possible source of the interference l Connect the equipmen...

Страница 20: ...separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interfer...

Отзывы: