Talley B.A.S.E RECLINER MAT Скачать руководство пользователя страница 13

12

OPERATIONAL CONDITIONS

A temperature range of +5 °C to +40 °C;
A relative humidity range of 15% to 93%, non-condensing; and
Operational Atmospheric Pressure: 700 hPa to 1060 hPa
Suitable for pollution degree 2

Operational altitude ≤ 2 000 m

IP Rating: IP21 pump only

TRANSPORTATION OF MATTRESS SYSTEM

The mattress should be loosely rolled lengthwise with the cover innermost, taking care not 
to strain the umbilical. It can then be transported and stored in the carry bag with the power 
unit, mains cable and this booklet. Do not stack bagged mattresses more than two high to 
avoid strain on the umbilical.

MANUFACTURER’S GUARANTEE

All power units, alternating mattresses/cushions and foam mattresses/cushions are covered 
by a 24 month manufacturer’s guarantee. The intended design life is 5 years if fully serviced.

Fault Finding

The power unit can be reset by pressing the MUTE button once.  This also silences the 
sounder and clears the message from the display screen.  All systems have a fault log that 
records the last 5 faults via the DATA display mode.  If problems reoccur contact Talley.

AC FAIL fault

 – indicates a mains power failure, a sounder will be heard if power is 

interrupted, e.g. power unit switched off, power cut, disconnection of mains lead. Press MUTE 
or re-connect to power supply.

ROTOR SYSTEM fault

 – indicates the automatic sequential cycle has stopped or there is a 

fault in the system.  Switch power off, press MUTE button, then switch power on again.  If the 
fault re-occurs, contact Talley.

LOW PRESSURE fault

 – indicates pressure has fallen below the minimum allowable levels.  

Check that the hose is connected to the power unit correctly.  Check that the CPR device is 
fully closed, i.e. the dial of the CPR device is rotated clockwise until it 

‘clicks’

 into the closed 

position    .  Check that the internal cells are connected and that no cell is punctured.  Press 
the MUTE button to clear the message and to silence the sounder. If the fault re-occurs, 
contact Talley.

EMI fault

 – indicates that the unit detects the pressure sensor amplifier is adversely affected 

by external RF fields. This will clear when interference ceases.

Содержание B.A.S.E RECLINER MAT

Страница 1: ...Pressure Area Care Mattresses and Cushions QUATTRO ACUTE Systems QUATTRO PLUS Systems QUATTRO OVERLAY Systems PULSAIR CHOICE Systems POLYFLOAT SUPREMA POLYFLOAT DORMIRA B A S E Seating Systems reduci...

Страница 2: ...the user to the possibility of a problem with the system associated with its use or misuse Operating Instructions IP Ingress Protection 2 Protection against fingers or other object not greater than 80...

Страница 3: ...n a cot size version for paediatric use l QUATTRO PLUS mattress replacement system for patients at high risk of pressure ulcers l QUATTRO OVERLAY mattress overlay system for patients at high risk of p...

Страница 4: ...sing or lowering of bed or bed rails or any other moving object l The power unit must only be used with a suitable approved cord and plug set as supplied by Talley l The system is not used in the pres...

Страница 5: ...injury l In order for alternating air pressure mattress cushion range to operate effectively please avoid placing objects on the surface that may obstruct the movement of air between the cells For th...

Страница 6: ...d mattress and causing possible injury to the patient A strap extension kit is available for use with divan beds part number 97 50 10 153 4 Suspend the power unit from the foot board of the bed having...

Страница 7: ...screen Press and hold the MUTE UNLOCK button until power unit beeps if further button operation is needed i e change of therapy mode or comfort setting 12 Place the carry bag and user manual in a safe...

Страница 8: ...The mattress and cushion will operate simultaneously allowing the simple transfer of the patient between the bed and chair NB When the patient is in bed return the comfort setting to the mattress set...

Страница 9: ...The power unit will lock again 2 minutes after the last button operation NB After power failure switching the power off pressing MUTE cancels the system s previous settings When power returns the def...

Страница 10: ...ain inflated so supporting the patient All alternating air pressure mattresses can be used on profiling bed frames slatted frames in filled frames and divans Back rests or pillows for support should b...

Страница 11: ...insed with clean water to remove any residues This will help prevent any possible long term compatibility issues associated with disinfectant residues l Alternatively disinfection may be achieved by l...

Страница 12: ...iversally used on all hard and soft surfaces without any detrimental effect TECcare CONTROL is CE marked for cleaning medical equipment SERVICING Once the initial guarantee period expires Talley recom...

Страница 13: ...en All systems have a fault log that records the last 5 faults via the DATA display mode If problems reoccur contact Talley AC FAIL fault indicates a mains power failure a sounder will be heard if pow...

Страница 14: ...by pressing MUTE button EMI FAULT Caused by external RF fields Fault will clear when interference ceases Check air cells and tubing are intact FAULT Reset power unit by pressing MUTE button MATTRESS...

Страница 15: ...gap of 2 5cm either side of the mattress should not be exceeded when side rails are deployed CUSHIONS Place the POLYFLOAT cushion onto a chair ensuring cushion is placed the correct way up N B It shou...

Страница 16: ...s will help prevent any possible long term compatibility issues associated with disinfectant residues l Alternatively disinfection may be achieved by laundering at temperatures not exceeding 650 C for...

Страница 17: ...Time QUATTRO ACUTE variable continuous QUATTRO PLUS 16 minutes continuous QUATTRO OVERLAY 16 minutes continuous PULSAIR CHOICE 16 minutes mattress replacement 8 minutes mattress overlay cushion conti...

Страница 18: ...30mm x 430mm x 70mm UK standard size B A S E RECLINER MAT cushion 1080mm x 430mm x 70mm PULSAIR CHOICE cushion 430mm x 430mm x 40mm Weight B A S E SEQUENTIAL cushion 1 9kg n B A S E RECLINER MAT cushi...

Страница 19: ...ering equipment repaired or replaced l Reorient or relocate the interfering equipment l Increase the separation between the equipment and the possible source of the interference l Connect the equipmen...

Страница 20: ...separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interfer...

Отзывы: