How to Calibrate
the Temperature
The
RX-802AS can be calibrated using a tip thermometer. Generally speaking, when
measuring the tip temperature of the
RX-802AS with a standard tip thermometer, the
measured temperature will usually be lower than that of the actual tip temperature
displayed on the
RX-802AS control unit display. This could be because the tip is not
coming into contact properly with the sensor of the tip thermometer. It could also be
because once the tip thermometer comes into contact with the
RX-802AS soldering
tip, the metallic parts in the thermometer (for example alumel chromel in the Type-K
thermocouple) conduct heat which transfers away from the
RX-802AS tip. This
results in the thermometer indicating a slightly lower temperature than the one in the
RX-802AS display.
Key Operation
TEMP Display
4-3 Calibration Function
3
(
Example
: Where the set temperature is 300°C (572°F),
tip thermometer value is 295°C (563°F), and the calibra-
tion value to be input is 5°C (9°F).)
Press the
key once.
Press the
key so that
5
(˚C)
is displayed.
Displays
alternately
Displays
alternately
Displays
alternately
Returns to the set
temperature display.
Setting completed
Press and hold the
key for over
3 seconds continuously.
Press and hold the
key for over
3 seconds continuously.
Calibration value is set to 0 before shipping. The tip number must be changed when the different type of
tip is mounted according to
4-2 Calibration of the Tip Temperature
.