9
2. Select Q.C. and press M to perform this test in the control
solution measuring mode.
Warning:
When doing the control solution test, you have to mark it so that the test
results will
NOT
mix with the lactate
TEST RESULTS
stored in the memory.
Failure to do so will mix up the lactate test results with the control solution
test results in memory.
3. Apply control solution (h)
Shake the control solution vial thoroughly before use. Squeeze
out the first drop and wipe it off, then squeeze out another drop
and place it on the tip of the vial cap. Hold the meter to move the
absorbent hole of the test strip to touch the drop. Once the
confirmation window fills completely, the meter will begin
counting down.
NOTE:
To avoid contaminating the control solution, do not directly apply control
solution onto a strip.
4. Read and compare the result
After counting down to 0, the control solution test result will
appear on the display. Compare this result with the range printed
on the test strip box or test strip vial and it should fall within this
range. If not, please read the instructions again and repeat the
control solution test.
NOTE:
The control solution range printed on the test strip box is for control solution
use only. It is not a recommended range for your lactate level.
See the
MAINTENANCE
section for important information about your
control solutions.