TACTILE
)
Problem
Recommended Solution
Cannot change
pressure
setting.
Once you start a treatment, settings cannot be changed
unless:
• The treatment program has completed .
• The treatment is paused .
• The ON/OFF button is pressed .
To change pressure setting while a therapy session is in
progress, first press the START/PAUSE button . Then settings
may be adjusted by pressing the PRESSURE button .
The garment
chambers do
not fill with air.
1 . Verify the ON/OFF and START/PAUSE buttons have been
pressed . (A green light will illuminate when the START/
PAUSE button is pressed .)
2 . Verify that the connectors are attached properly .
3 . Use the “Y” connector only when you are using both
garments .
4 . Ensure that both upper and lower connector latches are
engaged .
5 . If the chambers still do not fill, turn the machine off,
detach the connectors, and turn the machine ON again .
6 . Feel for air coming out of the controller unit, reconnect
the hose .
7 . Press the START/PAUSE button to begin a treatment
session .
Garment
pressures are
higher or lower
than expected.
1 . Verify that the correct pressure has been selected .
2 . Adjust the fit of the garment, ensuring no folds or kinks
are in the fabric .
3 . Check to be sure the latches are firmly in place and
attached properly .
Garment
remains
inflated.
It is normal for a small amount of air to remain in the
chambers between inflations, giving the garment a puffy
appearance . If the chambers remain fully inflated:
1 . Press the ON/OFF button twice in quick succession
(ON, OFF, ON, OFF) to allow for another deflation cycle .
2 . Ensure that the tubing is not kinked or pinched .
3 . Disconnect the connectors and the chambers should
deflate .