![TA DISCOVERY DSC2500 Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/ta/discovery-dsc2500/discovery-dsc2500_getting-started-manual_3112362041.webp)
Discovery DSC™ Getting Started Guide
Page 40
Installation of the LN Pump involves:
1
“Installing the Transfer Line” on page 40
2
“Installing the LN Pump to the Tank” on page 42
3
“Installing the LN Pump on the Discovery DSC” on page 43
Installing the Transfer Line
Before the LN Pump can be installed on the Discovery DSC, the transfer line must first be attached to the
LN Pump. Follow the instructions below to install the transfer line.
1
Loosen the two thumbscrews on either side of the vent cover. Pull the cover away and down from the
LN Pump to remove it.
Figure 20
Vent cover thumbscrews (left); Removing the vent cover (right).
2
Loosen the six thumbscrews on the head connection cover. Use a Phillips screwdriver, if needed. The
cover is removed in two pieces (top/front and sides/bottom/front). Lift the cover pieces up and away
from the LN Pump to remove it.
Figure 21
Head connection cover thumbscrews.