![TA DISCOVERY DSC2500 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/ta/discovery-dsc2500/discovery-dsc2500_getting-started-manual_3112362035.webp)
Discovery DSC™ Getting Started Guide
Page 34
4
Place the FACS over the cell, aligning the orientation hole in the FACS with the pin on the instrument
cell plate.
Figure 12
Placing the FACS over the cell.
5
Obtain a long 5/32-inch hexagonal (Allen) wrench from the accessory kit.
6
Insert the tip of the wrench into any one of the three captive screws in the FACS plate and begin to
tighten (3 to 4 turns). DO NOT fully tighten yet.
Figure 13
Securing the FACS to the cell (2 out of 3 screws shown).
7
Repeat
for the two remaining captive screws. After you have started each screw, go back and
tighten down all three screws (3 to 4 turns at a time) until you feel them touch the bottom. Do not over
tighten.
Captive screws
Alignment pin