Discovery DSC™ Getting Started Guide
Page 43
MISE EN GARDE: Un mauvais alignement de la pompe sur le réservoir peut endommager le
bouchon d'étanchéité sous vide. Pour éviter le risque de gel du capuchon d'étanchéité à vide, la fig-
ure 26 montre l'orientation de la pompe vers le réservoir.
Figure 26
Pump and tank orientation.
Installing the LN Pump on the Discovery DSC
Installation of the LN Pump onto the Discovery DSC involves:
1
Removing the Discovery DSC covers
2
Mounting the cooling head
3
Connecting the heater cable
4
Replacing the covers
5
Connecting the cables and lines
This section provides the instructions for installing the LN Pump on the Discovery DSC.
Remove the top cover on Discovery DSC 25, DSC 250, DSC 2500
1
If an Autosampler is present, it must be moved away from the cell by pressing the
Lid Open/Close
button on the user interface.
WARNING: Do not place your hands in the path of the Autosampler when it is in motion. Physi-
cal injury may occur.
AVERTISSEMENT: Ne placez pas vos doigts sur le chemin de l'échantillonneur automatique
lorsqu'il est en mouvement. Des blessures physiques peuvent se produire.
Return line
Supply line
Vacuum seal cap
Relief valves: DO NOT ALIGN
OVER VACUUM SEAL CAP
Vent: DO NOT ALIGN
OVER VACUUM SEAL CAP