NB/NHB Series Technical Manual
- 18 -
8. CALIBRATION
Turn on the scale.
Press
key during self checking, display will be show
f 1 unt
Press
key until display will be show
tech
Note: Before enter the tech menu, press calibration switch, which is
locating below the scale
Press
and
keys, display will be show
p 1 lin
8.1. Linear Calibration
li near
Press
key to enter calibration, display will be shown
pi n
Press
and
keys, display will be
Display will be shown
load 0
Ensure the platform is empty
After stable and zero indicator on, display will be shown
L oad1
Place the calibration weight (1/3 of the capacity) on the platform.
After stable and zero indicator on, display will be shown
L oad 2
Place the calibration weight (2/3 of the capacity) on the platform.
After stable and zero indicator on, display will be shown
L oad3
Place the calibration weight (full of the capacity) on the platform.
After stable and zero indicator on, display will be will be start self checking.
After self checking display will come to normal weighing mode
8.2. Normal calibration
p2 cal
Press
key to enter calibration, display will be show
unload
Remove all the weight from the platform.
When scale get stable, display will be show
load
Place the calibration weight (1/3 of the capacity or full of the capacity) on the
platform
After stable, scale will be calibrate automatically and will start self-test.
Note: Incase display will show any error message or incorrect measurement,
repeat the calibration again.