IHB++ Series Technical Manual
- 16 -
After stable and zero indicator on, display will be show
Load 2
Remove the third calibration mass weight from the pan
After stable and zero indicator on, display will be show
Load1
Remove the second calibration mass weight from the pan
After stable and zero indicator on, display will be show
load 0
Remove the
first calibration mass weight from the pan (
*Note: Ensure the pan is empty)
After stable and zero indicator on, display will be will be start self checking.
After self checking display will come to normal weighing mode
*Note:
During the linear calibration steps LOAD 1 to LOAD 4, mass weights want to
add on the pan. Don’t remove all mass weights for the next step.
During the linear calibration steps LOAD 4 to LOAD 1, mass weights want to
reduce from the pan. Don’t remove all mass weights for the next step.
8.2 Normal calibration
p2 cal
Press
key to enter calibration, display will be show kg or lb, press
key
to select the calibration unit kilograms or pounds, press
key ,display will be
show un
Load
.
Remove all the weight from the platform.
When scale get stable,display will be show
load
Place the calibration weight (1/3 of the capacity) on the platform
After stable, scale will be calibrate automatically and will start self-test.
Note: Incase display will show any error message or incorrect measurement,
repeat the calibration again.
8.3 Manual calibration by built-in weight
In the weighing mode, press and hold
key 3 seconds. The display will be
show
XCALX
, after 20 seconds, the display will be show
PASS
and come back
to the weighing mode.
8.4 Check internal counts
F3 CNT