6.2.2
Flow speed (S)
This slider allows to change the sample injection speed into the flow cuvette. Low speed
values result in a better precision and accuracy. A higher speed can be used when the
particle concentration is measured and accuracy is of lower relevance.
6.2.3
Light source (L)
This option allows to switch on/off the light sources.
Please notice Laser needs a few minutes before active.
6.3
1D/2D Plot properties
6.3.1 Plot name
A default plot name (H
x
, P
x
) is defined in the file.
Default names of the parameters will be displayed.
The user can modify for each plot the parameter
name (line
comment)
enter a name better matching
the experiment.
6.3.2 X Log On/Y Log On
This option will allow the user to change the plot’s
scaling. Note that changing the scale will require
you to adapt the gains of the PMTs.
6.3.3 Erosion levels
The erosion level will set a threshold on the data displayed (not the acquired data). Some
of the low frequency points will not be displayed allowing a better visual discrimination of
the higher frequency data (signal against background).
To obtain the properties of a plot area,
press the
Ctrl
+ right click on the plot.
6.3.4 X/Y Channel
The displayed parameters can be chosen from the drop down menu; giving the user the
list of the activated parameters.