Preoperative preparation
Required set
01.306.000 Pelvic C-Clamp II Set
The complete sterile pelvic C-clamp set should be kept
ready for use in the resuscitation room. Depending on the
type of injury, the orientation points in the pelvic region
of the injured person may be unclear.
Use an image intensifier during application of the pelvic
C-clamp to verify anatomic references.
– Anteroposterior (AP) plain pelvic radiograph if necessary,
oblique views (inlet and outlet) or CT
– Patient positioning must allow for intraoperative
fluoroscopic controls in AP, inlet and outlet projections
– Have an image intensifier available
Positioning
– Place the patient in a supine position
– To facilitate reduction, ensure free draping of the leg on
the injured side and good coverage of the genital region
– To prepare the patient, disinfect the proximal femur and
the buttocks and cover with sterile sheets
Preoperative Preparation