5
Apply pelvic C-clamp
Instruments
03.306.010 Pelvic C-Clamp II
03.306.011 8 mm Protective Cap
321.20
Ratchet Wrench, 11 mm width across flats
391.93
Wire Cutter, 220 mm
Be sure the fracture is correctly reduced before placement
of the pelvic C-clamp. After inserting the Kirschner wire on
the uninjured side, slide the clamp with cannulated nails
over the wire and ensure that the tip of the nail grips the
bone securely. Then place the second nail on the injured
side (no Kirschner wire is necessary on this side).
Note:
In cases of severe dislocations of the pelvis, pulling
on the leg, internal rotation and even lateral compression
may improve reduction and facilitate application of the
pelvic C-clamp.
Apply Pelvic C-Clamp
10 DePuy Synthes Pelvic C-Clamp II Surgical Technique