FIFO DATA PACKETS
Symmetricom Inc
bc635VME/350VXI Time and Frequency Processor (Rev. E)
4-17
Table 4-2
Time1 and Time2 Fields
Bit #
15-12
11-8
7-4
3-0
TIME0 Field
Not Defined.
Not Defined.
Status.
Unused.
TIME1 Field
2
28
Seconds.
2
24
Seconds.
2
20
Seconds.
2
16
Seconds.
TIME2 Field
2
12
Seconds.
2
8
Seconds.
2
4
Seconds.
Seconds.
TIME3 Field
10E-1 Seconds.
10E-2 Seconds.
10E-3 Seconds.
10E-4 Seconds.
TIME4 Field
10E-5 Seconds.
10E-6 Seconds.
10E-7 Seconds.
Not Defined.
4.1.15 PACKET “Q” SET DISCIPLINING GAIN
This packet allows the gain and sense of the disciplining process to be set via the host bus.
Originally this feature was used for Symmetricom developmental purposes, but it would also be
indispensable to anyone attempting to discipline an external oscillator using the TFP. The
format is as follows.
byte
1
SOH
byte
2
“Q”
byte
3
“0” - “F”
least significant nibble
byte
4
“0” - “F”
most significant nibble
byte
5
sense: “1” = positive, “0” = negative
byte
6
ETB
4.1.16 PACKET “S” SET YEAR
This packet allows users to set the year in Modes 0, 1, and 2. This is necessary to get the leap
year calculator to function in these modes. After writing the year you must wait at least one full
second before reading it back using the “O” packet.
byte
1
SOH
byte
2
“S”
byte
3
years tens
byte
4
years units
byte
5
ETB