AMU Powercon
Instrument Setup
A-96.250.331 / 230320
23
3.
Instrument Setup
3.1. Establish sample flow
1
Open flow regulating valve.
2
Wait until the flow cell has been completely filled.
3
Switch on power.
4
Adjust sample flow.
5
Let the instrument run-in for 1 h.
This recommendation is valid for rinsed cation exchanger
resin (nuclear grade) delivered by Swan.
Not rinsed cation exchanger resin from other suppliers may
require a run-in period of several hours to several days.
3.2. Programming
Sensor
parameters
Program all sensor parameters in Menu 5.1.2.1,
<Installation> <Sensors> <Sensor parameters>:
The sensor characteristics are printed on the label of each sensor.
Enter the:
Cell constant [cm
-1
]
Temperature correction [°C]
Cable length
Temperature compensation
87-344.203
UP-Con1000SL
Sensor type
SW-xx-xx-xx ZK = 0.0417
Cell constant
SWAN AG
DT = 0.06 °C
Temperature correction