AMU Powercon
Product Description
A-96.250.331 / 230320
7
Measuring
range
Range
0.005–0.999 µS/cm
1.00–9.99 µS/cm
10.0–99.9 µS/cm
100–999 µS/cm
1.00–2.99 mS/cm
3.0–9.9 mS/cm
10–30 mS/cm
Resolution
0.001 µS/cm
0.01 µS/cm
0.1 µS/cm
1 µS/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm
Automatic range switching. Values for a cell constant of
0.0415 cm
-1
(Swansensor UP-Con1000).
Accuracy ±1% of measured value ±1 digit.
Sensor cell
constant
0.005–10 cm
-1
Temperature
compensation
none
Coefficient
Neutral salts
High-purity water
Strong acids
Strong bases
Ammonia, Eth. am.
Morpholine
Temperature
measurement
Pt1000 type sensor (DIN class A)
Measuring range:
Resolution:
-30 to +250 °C
0.1 °C
Sample flow
measurment
with digital SWAN sample flow sensor