64
A-96.250.611 / 120617
AMI INSPECTOR Conductivity
Default Values
Relay 1 and 2
Function: .......................................................................... limit upper
Parameter: .................................................................... Conductivity
Setpoint:................................................................................. 30 mS
Hysteresis: .............................................................................. 10 µS
Delay:.....................................................................................30 Sec
If Function = Control upw. or dnw:
Parameter: .................................................................... Conductivity
Settings: Actuator: ...........................................................Frequency
Settings: Pulse Frequency: ............................................ 120/min.
Settings: Control Parameters: Setpoint: ............................. 30 mS
Settings: Control Parameters: P-band:................................ 10 µS
Settings: Control Parameters: P-band:................................. 1 mS
Settings: Control Parameters: Reset time: ...........................0 Sec
Settings: Control Parameters: Derivative Time: ...................0 Sec
Settings: Control Parameters: Control Timeout:................... 0 Min
Settings: Actuator:................................................. Time proportional
Cycle time: ............................................................................. 60 s
Response time: ..................................................................... 10 s
Settings: Actuator.......................................................... Motor valve
Run time: ............................................................................... 60 s
Neutral zone: ...........................................................................5%
If Function = Timer:
Mode:.................................................................................... Interval
Interval:................................................................................. 1 min
Mode: ....................................................................................... daily
Start time: ........................................................................00.00.00
Mode:.....................................................................................weekly
Calendar; Start time: .......................................................00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ...............................................................................10 Sec
Delay:.......................................................................................5 Sec
Signal output:............................................................................. cont
Output/Control: .......................................................................... cont