8
A-96.250.611 / 120617
AMI INSPECTOR Conductivity
Product Description
Special
Features
Many temperature compensation curves for specific conductivity
measurement:
none
Coefficient
Neutral salts
High-purity water
Strong acids
Strong bases
Ammonia, Eth. am.
Morpholine
Signal Output
(optional)
One signal output programmable for measured values (freely
scaleable, linear or bilinear) or as continuous control output (control
parameters programmable).
Current loop:
0/4–20 mA
Maximal burden:
510
Relay
Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function.
Maximum load:
100 mA/50 VAC
Alarm Relay
One potential free contact.
Alternatively:
Open during normal operation, closed on error and loss of
power.
Closed during normal operation, open on error and loss of
power.
Summary alarm indication for programmable alarm values and in-
strument faults.
Input
For potential-free contact to freeze the measuring value or to inter-
rupt control in automated installations (hold function or remote-off)
Safety features
No data loss after power failure, all data is saved in non-volatile
memory. Over voltage protection of in- and outputs.
Galvanic separation of measuring inputs and signal outputs.
Sensor
Swansensor UP-Con1000 SL high precision two-wire electrode
made of stainless steel with integrated NTC temperature probe.