99
SV 200A User Manual
Automatic system check
When the
Automatic system check
feature is enabled the calibration factor of the instrument will be
periodically verified using built-in electrostatic actuator. You can specify the hour and a weekday of the
automatic system checks.
To activate the automatic system check,
select
Enabled (no Cal. Factor update)
option from the
Mode
dropdown list.
To schedule automatic system checks,
choose weekday and select time of the system check from the
Do system check every day at (hh:mm)
dropdown list.
Note: System check cannot be considered as a calibration. Calibration factor will not be
updated during Automatic System Check procedure.
Note: Measurements are not stopped for the Automatic System Check procedure.
Calibration logging
Results of calibration (both manual and automatic) and automatic system check are stored in the
c
alibration and system check history file on the instrument’s SD-card. Log file name is
C.TXT
. The file
can be downloaded to the PC and then opened in any text editor, such as MS Notepad. Maximum
allowed size of log file is specified in the
Maximum calibration log file size [MB]
position.
9.8.12 Alarm settings
In the
Alarm configuration
section, press
Advanced Alarm Configuration
button to configure alarm
functionality of SV 200A. In the
Advanced Alarm Configuration
window, you can define and configure
the set of alarm
Conditions
and
Events
.
Alarm Conditions and
Events tabs
List of Alarm Conditions
Threshold conditions
configuration panel
Содержание SV 200A
Страница 123: ...123 SV 200A User Manual Appendix A REMOTE CONTROL firmware revision 1 x x Under development ...
Страница 153: ...153 SV 200A User Manual Free Field Frequency response of SV 200A ...
Страница 164: ...164 SV 200A User Manual 250 Hz 315 Hz 90 deg 0 deg 90 deg 0 deg 400 Hz 500 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 165: ...165 SV 200A User Manual 630 Hz 800 Hz 90 deg 0 deg 90 deg 0 deg 1000 Hz 1250 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 166: ...166 SV 200A User Manual 1600 Hz 2000 Hz 90 deg 0 deg 90 deg 0 deg 2240 Hz 2500 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 167: ...167 SV 200A User Manual 2800 Hz 3150 Hz 90 deg 0 deg 90 deg 0 deg 3550 Hz 4000 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 168: ...168 SV 200A User Manual 4500 Hz 5000 Hz 90 deg 0 deg 90 deg 0 deg 5600 Hz 6300 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 169: ...169 SV 200A User Manual 7100 Hz 8000 Hz 90 deg 0 deg 90 deg 0 deg 8500 Hz 9000 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 170: ...170 SV 200A User Manual 9500 Hz 10000 Hz 90 deg 0 deg 90 deg 0 deg 10600 Hz 11200 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 177: ...177 SV 200A User Manual 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 178: ...178 SV 200A User Manual 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 179: ...179 SV 200A User Manual 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...