8.4.1 Configuration views
The
Configuration
view consists of several tabs that enable configuring measurement parameters
(
MEASUREMENT SETUP
), data saving (
STORAGE
), export of measurement data into CSV files
(
CSV EXPORT
), audio recording (
AUDIO RECORDING
), events triggering (
EVENT TRIGGERS
),
calibration of the instrument (
CALIBRATION
), auxiliary settings (
AUXILIARY SETTINGS
) and
performing the firmware upgrade (
FIRMWARE UPGRADE
).
To load the new configuration to the station, click the
button.
MEASUREMENT SETUP tab
In the
MEASUREMENT SETUP
tab, you can:
1. Select the
Measurement function
:
Level Meter
,
Octave 1/1
,
Octave 1/3
2. Update the
Instrument clock
3. Select the type of
RMS/Leq Integration
:
Linear
or
Exponential
4. Select
Filter
(
Z
,
A
,
C
) and
Detector
type (
Impulse
,
Fast
,
Slow
) for profiles
5. Select
Filter
for the spectrum (position appears when the
Octave 1/1
or
Octave 1/3
function is
selected).
RMS/Leq Integration
defines the detector type for the calculations of the
Leq
,
LEPd
,
Lnn
and
SEL
measurement
functions.
Linear
integration is required when you wish to obtain the true RMS value of
the measured signal. When this option is selected the values of the
Leq
,
LEPd
,
Lnn
and
SEL
functions
do not depend on the detector time constant (
Fast
,
Slow
or
Impulse
), defined for the profiles.
Exponential
integration
is required in some standards for
Leq
measurements. When this option is
selected the values of the
Leq
,
LEPd
,
Lnn
and
SEL
measurement
results depend on the detector time
constant (
Fast
,
Slow
or
Impulse
), defined for the profiles.
Such measurement results like
Lmax
,
Lmin
,
Ltm3
or
Ltm5
are always calculated with the
Exponential
integration and selected time constants. And vice versa, such results as
Lpeak
don’t use integration
at all.
1
5
4
3
2
Содержание SV 200A
Страница 123: ...123 SV 200A User Manual Appendix A REMOTE CONTROL firmware revision 1 x x Under development ...
Страница 153: ...153 SV 200A User Manual Free Field Frequency response of SV 200A ...
Страница 164: ...164 SV 200A User Manual 250 Hz 315 Hz 90 deg 0 deg 90 deg 0 deg 400 Hz 500 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 165: ...165 SV 200A User Manual 630 Hz 800 Hz 90 deg 0 deg 90 deg 0 deg 1000 Hz 1250 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 166: ...166 SV 200A User Manual 1600 Hz 2000 Hz 90 deg 0 deg 90 deg 0 deg 2240 Hz 2500 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 167: ...167 SV 200A User Manual 2800 Hz 3150 Hz 90 deg 0 deg 90 deg 0 deg 3550 Hz 4000 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 168: ...168 SV 200A User Manual 4500 Hz 5000 Hz 90 deg 0 deg 90 deg 0 deg 5600 Hz 6300 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 169: ...169 SV 200A User Manual 7100 Hz 8000 Hz 90 deg 0 deg 90 deg 0 deg 8500 Hz 9000 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 170: ...170 SV 200A User Manual 9500 Hz 10000 Hz 90 deg 0 deg 90 deg 0 deg 10600 Hz 11200 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 177: ...177 SV 200A User Manual 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 178: ...178 SV 200A User Manual 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 179: ...179 SV 200A User Manual 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...